兔颅脑低剂量CT扫描的图像质量评价及辐射防护意义 被引量:9

The Evaluation of Radioprotection with Low Dose CT Scanning in Normal Rabbits Brain
下载PDF
导出
摘要 目的研究兔颅脑扫描中降低管电流及改变螺距与图像质量的关系,探讨低剂量扫描的可行性及其辐射防护价值。方法将15只新西兰兔分别按以下6组条件行颅脑CT检查:200mAs、螺距1.0组;200mAs、螺距1.5组;110mAs、螺距1.0组;110mAs、螺距1.5组;70mAs、螺距1.0组;70mAs、螺距1.5组。由两名有经验的放射科医师以双盲的方法使用4分制评分,对6组图像中头部6个部位的解剖结构的清晰度进行质量评分,比较各组间图像质量及CT剂量加权指数(weightedCTdoseindex,CTDIW)的差异性。结果所有图像质量均能满足诊断要求。两名阅读者对6组图像中6个解剖部位的总评分差异均无统计学意义(P>0.05),对每个解剖部位的评分中,6组图像两两比较,差异均无统计学意义(均P>0.05)。而70mAs、螺距1.5组较200mAs、螺距1.0组的CTDIW降低约76.7%。结论应用70mAs低剂量颅脑CT扫描是可行的,图像质量能够满足临床要求,而辐射剂量大幅下降。 Objective To determine wheather a lower radiation dose technique and various pitch could be used in CT of the rabbits' brain without jeopardizing the diagnostic accuracy of the images, and determine the evaluation of radioprotection with low dose CT scanning. Methods Fifteen rabbits underwent CT using 200 mAs,110 mAs or 70 mAs,and pitch 1.0 or 1.5. Anatomy details and the confidence level in reaching a diagnosis were evaluated by two radiologists in a double-blinded manner using a 4-point scoring system. The CTDIw of every group were compared. Results For both reader there was no statistically significant difference between 6 group total score of 1-6 anatomical detail and each of 6 anatomical detail although score for each of 6 anatomical detail. The CTDIw of 70 mAs,in pitch 1.5 group decreased about 76.7%. Conclusion Radiation dose reduction in brain CT is feasible in clinical use, and quality of images can be reserved. It plays an important role in radiation protection.
出处 《苏州大学学报(医学版)》 CAS 北大核心 2008年第2期215-217,共3页 Suzhou University Journal of Medical Science
基金 南通市科技局资助项目(S5822)
关键词 低剂量CT 头颅 体层摄影术 X射线计算机 辐射防护 low-dose CT rabbits brain tomography X-ray computed radioprotection
  • 相关文献

参考文献8

  • 1Calzdo A,Rodriguez R,Munoz A,et al. Quality criteria implementation for brain and lumber spine CT examinations[J]. Bri J Radiol,2000,73(868):384-395.
  • 2Chan CY,Wong YC,Chau LF,et al.Radiation dose reduction in pediatrics cranial CT [J ]. Pediatric Radial, 1999,29(10) :770-775.
  • 3Kaneko M,Eguchi K,Ohmatsu H,et al.Peripheral lung cancer:screening and detection with low-dose spiral CT versus radiography[J].Radiology, 1996,201 (3):798-802.
  • 4Haaga JR.Radiation dose management:weighing risk versus benefit[J]. AJR, 2001,177(2):289-291.
  • 5Kalra MK,Maher MM,Toth TL,et al.Strategies for CT radiation dos optmization[J].Radiology, 2004,230(3): 619- 628.
  • 6Mathlias Cohnen,Hubert Fischer,JUrgen Hamacher,et al.CT of the head by use of reduced current and kilovoltage:relationship between image quality and dose reduction[J]. AJNR, 2000,21 (10): 1654-1660.
  • 7陈伟,李文政,唐友林.螺旋CT图像噪声的评价[J].放射学实践,2003,18(7):535-536. 被引量:9
  • 8赵斗贵,郭文臣.延长CT球管寿命的经验[J].中国医学影像学杂志,2002,10(5):370-370. 被引量:2

二级参考文献5

  • 1王鸣鹏,吴威岚,张国桢.螺旋CT扫描成像质量分析[J].中华放射学杂志,1996,30(6):373-376. 被引量:14
  • 2Crawford CR,King KF.Computed tomography scanning with simultaneous patient translation[J].Med Phys,1990,17 (6):967-982.
  • 3Polacin A,Kalender WA,Marchal G.Evaluation of section sensitivity profiles and image noise in spiral CT[J].Radiology,1992,185(1):29-35.
  • 4Wang G,Vannier MW.Helical CT image noise-analytical results[J].Med Phys,1993,20(6):1635-1640.
  • 5Wang G,Vannier MW.Longitudinal resolution in volumetric X-ray computerized tomography-analytical comparison between conventional and helical computed tomography[J].Med Phys,1994,21(3):429-433.

共引文献9

同被引文献81

引证文献9

二级引证文献115

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部