血铅同位素稀释质谱法检测结果的传递性检验 被引量:2

Verification for Blood Lead Result of Magnitude Transfer Detected by Isotope Dilution-mass Spectrometry
下载PDF
导出
摘要 目的用原子吸收光谱法对同位素稀释-质谱方法检测牛血铅数据作量值传递性验证。方法原则采用卫生部推荐的原子吸收光谱法。首先通过检测表达中国有证标准物质,以示方法的可重复性和可操作性,在此基础上,用美国国家标准与技术研究院(NIST)研制的有证标准物质SRM955b的血铅浓度值作为确定数值制作工作曲线,检测专门研制的牛血铅物质,并与同位素稀释-质谱检测数据进行比对分析。结果两种方法所测三种浓度结果分别进行配对t检验,差异无显著性(P>0.05)。结论通过比对分析,原子吸收光谱法在严格质量控制条件下可以进行权威检测方法的量值传递。 Objective To verify bovine blood lead as magnitude transfer detected by atomic absorption spectroscopy method with isotope dilution-mass spectrometry. Methods Principles recommended by the Ministry of Health AAS method. Firstly detected certification standards material ,to show the way repeatability and operability. With the United States Institute of Standards and Technology (NIST) developed a certification standard material SRM955b blood lead concentration values defined numerical production curve. Detected bovine blood compared data with isotope dilution-mass spectrometry. Results Three concentration results measured by the two methods,respectively paired t test,there was no significant difference (P〉0.05). Conclusion Through ratio analysis,method of atomic absorption spectrometry in strict quality control conditions competent authority detection of magnitude transfer.
出处 《现代检验医学杂志》 CAS 2008年第3期37-39,共3页 Journal of Modern Laboratory Medicine
基金 2006年西安市科技攻关计划-社会发展攻关项目(GG06159)
关键词 同位素稀释质谱法 原子吸收光谱法 血铅 量值传递 isotope dilution mass spectrometry atomic absorption spectrometry blood lead magnitude transfer
  • 相关文献

参考文献8

  • 1Heumann Klaus G. Isotope dilution mass spectrometry. In Adayms F,Gijbels R and Van Grieken R. 祝大昌译.无机质谱法[M].上海:复旦大学出版社,1993:301.
  • 2刘炳寰.质谱学方法与同位素分析[M].北京:科学出版社,1993.
  • 3王仕俊.质谱学极其在核科学技术中的应用[M].北京:原子能出版社,1998.
  • 4BIPM. Commite consultative Pour La Quantite de Matiere[S].1996: 2,38.
  • 5Wattere Jr RL,Eberhardt KR,Beary ES,et al. Isotope Dilution Using Inductively Ceupled Plasma-Mass Spectrometry (ICP-MS) as a Primary Method [J]. Metro-lgia, 1997,34 : 87.
  • 6National Institute of Standards & Technology Certificate of Analysis. Standard Reference Material 955b Lead in Bovine Blood[S]. 2004:1-24.
  • 7In vitro diagnostic medical devices-Measurement of quantities in biological samples-Metrological traceability of values assigned to calibrators and control materials [S]. International Standard ISO17511. 2003,4. 2.2-4. 2.4.
  • 8陈文祥,申子瑜,郭健,杨振华.临床检验量值溯源中的重要术语与概念及有关问题[J].中华检验医学杂志,2005,28(2):142-146. 被引量:35

二级参考文献9

  • 1International Organization for Standardization. In Vitro Diagnostic Medical Devices- Measurement of Quantities in Samples of Biological Origin-Description of Reference Materials. ISO 15194, 2003.
  • 2International Organization for Standardization. Clinical Laboratory Medicine-Requirements for Reference Measurement Laboratories. 1SO 15195,2003.
  • 3International Organization for Standardization. General Requirements for the Competence of Testing and Calibration Laboratories. ISO 17025, 1999.
  • 4International Organization for Standardization. Medical Laboratories-Particular Requirements for Quality and Competence. ISO 15189, 2003.
  • 5Dybkaer R. Vocabulary for use in measurement procedures and description of reference materials in laboratory medicine. Eur J Clin Chem Clin Biochem ,1997, 35 : 141-173.
  • 6JJF1001—1998.通用计量术语及定义.中华人民共和国国家计量技术规范.[S].国家质量技术监督局,..
  • 7International Organization for Standardization. In Vitro Diagnostic Medical Devices-Measurement of Quantities in Samples of Biological Origin -Metrological Traceability of Values Assigned to Calibrators and Control Materials. ISO 17511. 2003.
  • 8International Organization for Standardization. In Vitro Diagnostic Medical Devices- Measurement of Quantifies in Samples of Biological Origin-Metrological Traceability of Values Catalytic Concentration of Enzymes Assigned to Calibrators and Control Materials. 1SO 18153, 2003.
  • 9International Organization for Standardization. In Vitro Diagnostic Medical Devices- Measurement of Quantities in Samples of Biological Origin-Presentation of Reference Measurement Procedures. ISO 15193, 2003.

共引文献34

同被引文献43

  • 1Guengerich FP,Johnson WW,Shimada T,et al.Activation anddetoxication of aflatoxin B1[J].Mutat Res,1998,402 (1/2):121-128.
  • 2Guengerich FP,Arneson KO,Williams KM,et al.Reaction of aflatoxin B1 oxidation products with lysine[J].Chem Res Toxicol,2002,15 (6):780-792.
  • 3Poirier MC,Santella RM,Weston A.Carcinogen macromolecularadducts and their measurement[J].Carcinogenesis,2000,21 (3):353-359.
  • 4Wild CP,Turner PC.The toxicology of aflatoxins as a basis for public health decisions[J].Mutagenesis,2002,17(6):471-481.
  • 5Wang JS,Huang TR,Su JJ,et al.Hepatocellular carcinoma and aflatoxin exposure in Zhuqing village,Fusui county,People's Republic of China[J].Cancer Epidemiol Biomarkers Prey,2001,10 (2):143 -146.
  • 6Long XD,Yun M,Huang HD,et al.Polymorphism of XRCC1 and the frequency of mutation in Codon 249 of the p53 gene in hepatocellular carcinoma among Guangxi population,China[J].Molecular Carcinogenesis,2008,47 (4):295-300.
  • 7李玉林.病理学[M].3版.北京:人民卫生出版社,2003:408.
  • 8Zhang YJ,Yu C,Ahsan H,et al.Inactivation of the DNA repair gene 06-methylguanine-DNA methyltransferase by promoter hypermethylation and its relationship to aflatoxin B1 -DNA adducts and p53 mutation in hepatocellular carcinoma[J].Int J Cancer,2003,103(4):440-444.
  • 9Dash B,Gyawu EA,Huebner HJ,et al.Determinants of the variability of aflatoxin-albumin adduct levels in ghanaians[J].J Toxicol Environ Health,2007,70(1):58-66.
  • 10Wang JS,Qian GS,Zarba A,et al.Temporal patterns of aflatoxin-albumin adducts in hepatitis b surface antigen-positive and antigen-negative residents of Daxin,Qidong County,People's Republic of China[J].Cancer Epidemiol Biomarkers Prey,1996,5 (4):253-261.

引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部