期刊文献+

激光器精密恒温水循环系统的研制

Development of Precision Constant-temperature Water Circulator for Laser Generator
下载PDF
导出
摘要 目的:研制精密恒温水循环系统,以保证半导体激光器在恒定温度下正常工作。方法:采用PT100传感器、恒流源激励电路、V-F转换器、单片机和加热器等电路研制恒温循环水温度控制系统,运用PID控制原理,通过精心选择PID的各种参数,实现高精度的恒温控制。结果:恒温循环水温度控制系统温控的测量精度达到±0.05℃。结论:该恒温循环水温度控制系统可以保证半导体激光器的正常工作温度。 Objective To provide Laser generator with a precision constant-temperature environment. Methods This system was developed on the bases of PT100 sensor circuit, MCU and SSR. The inputting PT100 voltage was first amplified and deflected, and then delivered to AD650 to perform WF. The outputting frequency was measured with high precision by TO port of MCU C8051F020, and the micro-controller calculated the result and transmitted PID output signal by 16 bit PWM to heater.Results This precision constant-temperature water circulator was gifted with an error within ±0.05℃.Conclusion This precision constant-temperature water circulator can make an ideal temperature environment available for laser generator. [Chinese Medical Equipment Journal,2008,29(6) :60-61,63]
出处 《医疗卫生装备》 CAS 2008年第6期60-61,63,共3页 Chinese Medical Equipment Journal
基金 空军总医院青年资助课题
关键词 恒温 激光器 控制 constant temperature laser generator control
  • 相关文献

参考文献3

  • 1杨宣访,卜乐平.《舰船数字式多点测温装置改进研究》鉴定材料[R].北京:海军工程大学.2002.6-9.
  • 2吴新杰 关建新.铂电阻线性化测温电路原理分析.仪表技术与传感器,1999,(1):15-17.
  • 3李英顺,伦淑娴.模糊PID温度测控仪[J].仪表技术与传感器,2003(1):20-22. 被引量:34

二级参考文献3

  • 1李华.MCS-51系列单片机实用接口技术[M].北京:航空航天大学出版社,1996..
  • 2胡汉才.单片机原理及接口技术[M].北京:清华大学出版社,1999..
  • 3张麦德,徐时亮,赵时英MCS-51单片机实用子程序及其应用上海:复旦大学计算机科学系微机开发应用研究室,2000.