摘要
目的:建立职业病人尿液中锑的原子荧光光谱测定方法。方法:样品采用压力消解法进行处理,使用硫脲-抗坏血酸-碘化钾混合溶液作为预还原剂,研究了尿中锑原子荧光光谱测定方法的最佳测定条件。结果:本法选择20 g/LKBH4溶液作为还原剂,50 g/L硫脲-50 g/L抗坏血酸-50 g/L碘化钾混合溶液作为预还原剂,线性范围为0~10μg/L,相关系数r≥0.9999,锑检出限分别为0.027μg/L,加标回收率为94.7%,锑的相对标准偏差为1.69%。结论:该方法简便快速、灵敏度高、准确度好,适合尿液中锑的测定。
Objective:A robust, accurate and sensitive analytical procedure was developed for the determination of trace antimony in urine of occupational patients by hydride generation atomic fluorescence spectrometry (HG- AFS). Methods: A method had been described for the determination of stibium (Sb) in the best condition in urine by hydride generation - atomic fluorescence spectrometry. Results:Reducer was 20 g/L KBH4. r was 0. 9999 in the range of 0 ~ 10 μg/L The detection limit was 0. 27μg/L, recovery 94. 7 and RSD 1.69. Conclusion:Using this method to determine Sb in urines is simple, reliable and accurate.
出处
《中国卫生检验杂志》
CAS
2008年第6期1062-1063,共2页
Chinese Journal of Health Laboratory Technology
关键词
尿液
锑
原子荧光光谱
Urines
Antimony
Atomic fluorescence spectrometry