期刊文献+

智能型高精度防护板法导热仪的设计 被引量:3

Design of Intelligent and High Precision Thermal Conductivity Apparatus with Guarded Plate
下载PDF
导出
摘要 介绍了一种基于AT89C51的智能型防护板法导热仪,通过对传统导热仪的结构、电源、加热方式、控制电路进行改进,使仪器在系统稳定性、精确性方面得到了提高。由于采用了恒温与恒功率相结合的加热方法,使得该仪器在测试的速度和精度上都有所提升,同时也为基于稳态法的导热议的设计提供了一种新思路。 The intelligent thermal conductivity apparatus with guarded plate and based on AT89C51 is introduced. Through improving the structure, power supply, heating mode and control circuit of traditional apparatus, the performance of instrument has been enhanced in stability and reliability. Because the combined heating method of constant temperature and constant power is used, the measurement accuracy and measuring speed of the instrument are increased. It provides a new idea for designing thermal conductivity apparatus based on steady-state measurement.
出处 《自动化仪表》 CAS 2008年第6期56-58,共3页 Process Automation Instrumentation
关键词 AT89C51 导热仪 一维稳态传热 导热系数 PWM 防护板法 智能型 AT89C51 Thermal conductivity apparatus One-dimensional steady state heat conduction Heat conductivity coefficient PWM Method of guarded plate Intelligent
  • 相关文献

参考文献6

  • 1PONTE F, FILIPPO P. Design criteria for guarded hot plate apparatus [ M ]. USA : ASTM, 1974:97 - 117.
  • 2杨惠林,徐骏华.高准确度防护热板法导热仪[J].同济大学学报(自然科学版),1994,22(2):259-262. 被引量:8
  • 3陶文栓,杨世铭.传热学[M].3版.北京:高等教育出版社,1998:71-78.
  • 4丁元杰.单片微机原理及应用[M].北京:机械工业出版社,2002.
  • 5杨金岩,郑应强,张振仁.8051单片据传输接口扩展技术与应用实例[M].北京:人民邮电出版社,2005:188-193.
  • 6边远春,王志强.MCS51单片机应用程序开发[M].北京:人民邮电出版社,2005:396-397.

二级参考文献1

  • 1曹声含,1990年

共引文献12

同被引文献25

  • 1郝丽宏,林凌,李刚,刘楚翘.改进的绝热材料导热系数防护热板法测控系统的设计[J].仪器仪表学报,2003,24(z1):360-361. 被引量:8
  • 2李春辉,张金涛.测温孔对保护热板法测量导热率的影响[J].计量学报,2007,28(z1):115-117. 被引量:2
  • 3宋杰.人工巢箱的制作和使用[J].生物学通报,1994,29(4):25-27. 被引量:16
  • 4杨惠林,徐骏华.高准确度防护热板法导热仪[J].同济大学学报(自然科学版),1994,22(2):259-262. 被引量:8
  • 5吴江涛,潘江,张可,刘志刚.一种全自动的准稳态法导热系数测量装置[J].自动化仪表,2005,26(5):23-25. 被引量:10
  • 6Siu M C I, Bulik C. Line - heat - source guarded - hot - plate apparatus [ J ]. Review of Scientific Instruments, 1981, 52 (11): 1 709-1 716.
  • 7Hemminger W, Jugel R. A guarded hot - plate apparatus for thermal conductivity measurements over the temperature range -75 to 200℃ [ J ]. International Journal of Thermophysic, 1985, 6 (5): 483-498.
  • 8Zarr R R, Flyrm D R, Hettenhouser J W, et al. Fabrication of a guarded - hot - plate apparatus for use over an extended temperature range and in a controlled gas atmos- phere [ R ]. New Brunswick: Thermal Conductivity 28/ Thermal Expansion 16 Joint Conferences, 2006:235-245.
  • 9James Filla B. A steady - state high - temperature apparatus for measuring thermal conductivity of ceramics [J]. Review of Scientific Instruments, 1997, 68 (7) : 2 822 -2 829.
  • 10Flyrm D R, Healy W M, Zarr R R. High -temperature guarded hot plate apparatus - Optimal locations of circular heaters [ R ] New Brunswick: Thermal Conductivity 25/ Thermal Expansion 16 Joint Conferences, 2006:466-476.

引证文献3