期刊文献+

射频消融导管温度感应精度的影响因素

Factors Influencing the Temperature-Sensing Accuracy of Ablation Catheters
下载PDF
导出
摘要 从TC线插入消融电极TC孔的深度和TC孔的直径两方面分析了其对温度感应的影响;对比不同公司的产品温度感应精度的差异。 Factors influencing the temperature-sensing accuracy of an ablation catheter are analyzed, in this paper, from the two aspects of the thermocouple temperature sensor, which are the TC length and the TC hole's diameter of the ablation electrode.Meanwhile,differences betweeen products from different companies are given too.
出处 《中国医疗器械杂志》 CAS 2008年第4期249-252,共4页 Chinese Journal of Medical Instrumentation
关键词 心律失常 射频 消融导管 温度 热电偶 arrhythmias, radiofrequency, ablation catheter, temperature, thermocouple
  • 相关文献

参考文献12

  • 1陈良华,刘同宝.射频消融治疗快速性心律失常[EB/OL].http://zhidao.baidu.com/question/6509029.html,2006-04-26/2007-05-01
  • 2A+医学百科.心律失常[EB/OL].http://www.a-hospital.com/%D0%C4%C2%C9%CA%A7%B3%A3.html,2007-05-01/2007-05-01
  • 3郑强荪.导管射频消融术治疗快速心律失常[EB/OL].http://www.37c.com.cn/education/credithour/dgxrs/default.asp?filename=01/0102.2007-05-02/2007-05-02
  • 4高山.射频治疗仪的种类和特点简介[J].中国医疗器械信息,2006,12(8):32-34. 被引量:3
  • 5H Calkins, E Prystowsky, M Carlson, et al. Temperature monitoring during radiofrequency catheter ablation procedures using closed loop control[J].Circulation, 1994,90(3): 1249-1286
  • 6Hiroshi Nakagawa, William S. Yamanashi, Jan V. Pitha, et al. Comparison of In Vivo Tissue Temperature Profile and Lesion Geometry for Radiofrequency Ablation With a Saline-Irrigated Electrode Versus Temperature Control in a Canine Thigh Muscle Preparation[J].Circulation, 1995,91:2264-2273
  • 7刘晓霞冷盐水灌注导管在导管消融术中的应用[EB/OL].http://www.365heart.com/shownews.asp?id=14262,2006-09-22/2007-05-03
  • 8中华医学会心电生理和起搏分会心房颤动治疗专家组.心房颤动认识和建议2006(二)[EB/OL].http://www.dxy.cn/portal/article/2/14/419/1833_1.html,2006-09-23/2007-05-09
  • 9JJ Langberg, M Gallagher, SA Strickberger, et al. Temperatureguided radiofrequency catheter ablation with very large distal electrodes[J] .Circulation, 1993,88(1 ):245 -249
  • 10那开宪,余平.射频消融术的临床应用[J].首都医药,2006,13(20):31-33. 被引量:2

二级参考文献5

  • 1Cosio FG, lopez-Gil M. Goicolea A , et al. Radiofrequency ablation of the inferior Vena cava-Trecaspia Valve isthmas in common atrial flatter.Am J Cardiol,1993,71:705-709
  • 2Zipes DP, Jackman WN, Dimarco JP, etal.Guidelines for clinical intracardiac electrophy siological and catheter ablation procedures.J Am Coil Cardil 1995,26:555-573
  • 3Chen SA, Tai CT, Tsai CF, et al .Radiofrequency catheter ablation of atrial fibrillation initiated. By pulmonary vein ectopic beats.J CardioVase Electrophysiol 2000,11:218-227
  • 4Clague JR, Dagres N, Kott Kamp H, etal. Targeting the slow pathway far atrioventricular nodal reentrant tachy cardia Initial results and long term follow up in 379 consecutive patients.Eur Heart J 2001,22 (1):82-88
  • 5Scheuimman MM,Huany S. The 1998 NASPE prospective catheter ablation registry PACE .2000,23(6): 1020-1028

共引文献3