期刊文献+

反相流动注射催化光度法测定痕量锰的研究 被引量:1

Revers Flow Injection-catalytic Spectrophotometry for Determination of Trace Manhanese
下载PDF
导出
摘要 本文基于在氨三乙酸活化下,锰(Ⅱ)对高碘酸钾氧化罗丹明B褪色的催化作用,建立了痕量锰的反相流动注射分析(rFIA)新方法。该法选择罗丹明B溶液的最大吸收波长555nm作为测定波长,方法的线性范围为0~180μg/L,检出限为5.5μg/L锰(Ⅱ),对20μg/L锰平行测定11次的相对标准偏差为1.5%,进样频率为36次/h。方法应用于水样、黄豆和标准铝合金样中痕量锰的测定。 A rapid catalytic spectrophotometric method for the determination of trace manganese based on manganese catalyzing decoloration reaction of rhodamin B with potassium periodate in the presence of nitrilo triacetic acid has been established by reverse flow injection technique. The detecting wavelength was 555nm ,the linear range is 0_180ug/L and the detection limit is 5 5ug/L. The relative standard deviation is 1 5% obtained from 11 standard solution each containing20ug/L of manganese. 36 samples can be analysed per hour. The method was applied to the determination of manganese in nature water. Soybean and aluminium alloy with satisfactory results.
作者 刘秀萍
机构地区 晋中师专化学系
出处 《山西农业大学学报》 CAS 1997年第1期81-84,共4页
关键词 流动注射分析 催化光度法 罗丹明B Flow injection analysis, catalytic spectro-photometry,Rhodamine B,Manganese