IOL-Master在人工晶状体屈光度计算中的精确性研究 被引量:5

Accuracy of IOL-Master for power calculation of artifical intraocular lens
下载PDF
导出
摘要 目的评价IOL-Master在人工晶状体屈光度计算中的临床应用价值。方法对160眼白内障分别运用IOL-Master与A型超声波测量眼轴,术后1月、3月行屈光检查。结果IOL-Master测量眼轴:22<L≤24.5mm组平均轴长为(23.75±0.56)mm,24.5<L≤26mm组平均轴长为(25.21±0.61)mm,L>26.0mm组平均轴长为(28.87±0.58)mm。A型超声波测量眼轴,上述3个组平均轴长依次为:(23.68±0.81)mm,(25.01±0.90)mm,(29.33±1.02)mm。术后1月,L>26.0mm组,A超组和IOL-Master组的平均等效球镜屈光度误差为(1.449±0.723)D及(0.875±0.682)D(P<0.001)。术后3月,L>26.0mm组,A超组和IOL-Master组的平均等效球镜屈光度误差为(1.345±0.714)D及(0.865±0.683)D(P<0.001)。结论IOL-Master在人工晶状体屈光度计算方面具有极高的临床应用价值。 Objective To evaluste the clinical aplication and Accuracy of IOL-Master for Power Calculation of Artificial Intraocular Lens. Methods Data was analyzed from 160 eyes with age-related cataracts who underwent phacoemulsification and lens implantation in our hospital from Jan. 2007 to Fed. 2007. The length of ocular axis and keratometric power were measured preoperatively by IOL-Master,uhrasonic biometry and auto-keratometer. Their refraction outcome were followed-up 1-3 month after operation. Results The ocular axial length that measured by IOL-Master was ( 23.75 ± 0. 56) mm in 22 〈 L ≤ 24.5 mm group, ( 25.21 ± 0.61 ) mm in 24.5 〈L≤26 mm group,and (28.87 ±0.58)mm in L 〉26.0 mm group. The ocular axial length of the three groups that measured by A-scan ultrasonic biometry were (23.68 ± 0.81 ) mm, (25.01 ± 0.90) mm and (29.33 ± 1.02)mm respectively. One month after surgery,the mean absolute refractive error in L 〉 26.0 mm group was ( 1. 449 ± 0.723 ) D by A-scan, and was (0.875 ± 0. 682 ) D by IOL-Master ( P 〈 0. 001 ). That was ( 1. 345 ± 0. 714 ) D measured by A-scan and was (0. 865 ± 0. 683 ) D measured by IOL-Master (P 〈 0. 001 ) 3 months after surgery. Conclusions IOL-Master has high clinical value for calculating the power of artificial IOLs.
出处 《眼外伤职业眼病杂志》 北大核心 2008年第7期530-532,共3页 Journal of Injuries and Occupational Diseases of the Eye with Ophthalmic Surgeries
关键词 IOL-MASTER A超 晶状体 人工 IOL-Master A-scan Intraocular lens, artificial
  • 相关文献

参考文献10

  • 1Wunder H. Increase A-scan accuracy for improved outcomes. Review of Ophthalmology ,2003,10( 1 ) :36 -38.
  • 2Shammas HJ. Intraocular lens power calculation [ M ]. America: SLACK Incoporated ,2004,146.
  • 3Rajan MS, Keihom I, Bell JA. Partial coherence laser interferometry vs conventional ultrasound biometry in intraocular lens power calculations [ J]. Eye,2002,16 (5) :552 - 556.
  • 4Verhulst E, Vrijghem JC. Accuracy of intraocular lens power calculations using the Zeiss IOL-Master: a prospective study [ J]. Bullsoc Beige Ophtalmol,2001,28( 1 ) :61 -65.
  • 5Drexler WF, Menapace R. Improved predication of intraocular lens power using partial coherence interferometry[ J]. J Cataract Refract Surg,2001,27 (6) :861 - 867.
  • 6Packer M, Fine H, Hoffman RS. Immersion A-scan compared with partial coherence interferometry outcomes analysis[ J]. J Cataract Refract Surg,2002,28 (2) :239 - 242.
  • 7Olsen T. Sources of error in intraocular lens power calculations[ J]. J Cataract Refract Surg, 1992,18(5 ) : 125 - 129.
  • 8Clam A K,Clan R,Pang PCK. The repeat ability and accuracy of axial lenth and anterior chamber depth measurements form the IOL-Master TM [ J ]. Ophthalmic and Physiological Optics, 2001,21 ( 6 ) : 477 - 483.
  • 9Kielhom I, Madhavan S, Peter M, et al. Clinical assessment of the Zeiss IOL-Master[ J]. J Cataract Refract Surg,2003 ,29 :518 -522.
  • 10Haigis W, Lege B, Miller N, et al. Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis[ J]. Grade's Arch Clin Exp Ophthalmol, 2000,238(9) :765 -773.

同被引文献35

  • 1黄振平.屈光手术的最新技术及其临床应用[J].医学研究生学报,2007,20(4):337-338. 被引量:22
  • 2Meldrum LM,Aaberg TM,Pate A,et al.Cataract extraction after silicon oil repair of retinal detachments due to necrotizing retinitis.Arch Ophthalmol 1996;114(7):885-892.
  • 3Reddy AR, Pande MV, Finn P, et al. Comparative estimation of anterior chamber depth by altrasonography, Orhsean Ⅱ,and IOL Master.J Cataract Refract Surg 2004 ; 30 ( 6 ) : 1268-1271.
  • 4Nemeth J, Fekete O, Pesztenlehrer N. Optical and ultrasound measurement of axial length and anterior chamber depth for intraocular lens power calculation.J Cataract Refract Surg 2003 ;29( 1 ) :85-88.
  • 5祁媛媛,张丰菊.光学相干生物测量仪的临床应用[J].中国实用眼科杂志,2007,25(8):809-812. 被引量:9
  • 6马楠,卢海,杨文利.声速调整法在硅油填充眼眼轴测量中的应用[J].眼科研究,2007,25(9):697-699. 被引量:12
  • 7李凤鸣.中华眼科学[M].北京:人民卫生出版社,2005.2467-2473.
  • 8Olsen T. Sources of error in intraocular lens power calculation [ J]. J Cataract Refract Surg, 1992,18 (2) : 125 - 129.
  • 9Shammas H. Intraocular lens power caculation [ M ]. America: SLACK Incorporated,2004 : 161.
  • 10Findl O, Kriechbaum K, Sacu S, et al. Influence of operator experience on the performance of ultrasound biometry compared to optical biometry before cataract surgery [ J ]. J Cataract Refract Surg ,2003,29 ( 10 ) : 1 950 - 1 955.

引证文献5

二级引证文献12

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部