摘要 目的以聚合酶链反应(PCR)检测mecA基因作为标准方法,评价Vitek-32微生物分析仪检测耐甲氧西林金黄色葡萄球菌(MRSA)及凝固酶阴性葡萄球菌(MRCNS)的可靠性。方法收集金黄色葡萄球菌72株,MRCNS64株,采用微生物分析仪Vitek-32配套的GPS-110卡检测MRSA与MRCNS,并与PCR及头孢西丁纸片扩散法进行比较。结果Vitek-32微生物分析仪检测MRSA及MRCNS的灵敏度及特异性分别为100.0%和89.7%、100.0%和80.0%;头孢西丁纸片扩散法的灵敏度及特异性均为100.0%。结论微生物分析仪Vitek-32操作简单、快速,检测MRSA及MRCNS的灵敏度高,是一种较好的方法。对于可疑株,用头孢西丁纸片扩散法复查可提高特异性。
作者 罗湘蓉 陈燕
出处 《检验医学与临床》 CAS 2008年第16期995-996,共2页 Laboratory Medicine and Clinic
  • 相关文献

参考文献3

  • 1熊艳,刘莹,吴正学,曾吉,杨汉菊.耐甲氧西林金黄色葡萄球菌3种检测方法的评价[J].中国卫生检验杂志,2005,15(2):209-210. 被引量:6
  • 2[3]Louie L,Matsumura SO,Choi E,et al.Evaluation of three rapid methods for detection of methicillin resistance in Staphylococcus aureus[J].J Clin Microbiol,2000,38(6):2170-2173.
  • 3[4]Felten A,Grandry B,Lagrange PH,et al.Evaluation of three techniques for detection of low-level methicillin-resistant staphylococcus aureus (MRSA):a disk diffusion method with cefoxitin and moxalactam,the Vitek 2 system,and the MRSA-screen latex agglutination Test[J].J Clin Microbiol,2002,40(10):2766-2771.

二级参考文献9

  • 1Sakoulas G,Gold HS,Venkataraman L,et al.Methicillin-resistant Staphylococcus aureus:comparison of susceptibility testing methods and analysis of mecA-positive susceptible strains[J].J Clin Microbiol,2001,39(11):3946-3951.
  • 2Hiramatsu K,Cui L,Kuroda M,et al.The emergence and evolution of methicillin-resistant Staphylococcus aureus[J].Trends Microbiol,2001,9(10):486-493.
  • 3Chambers HF.Methicillin resistance in staphylococci:molecular and biochemical basis and clinical implications[J].Clin Microbiol Rev,1997,10(4):781-791.
  • 4Zone diameter interpretive standards and equivalent minimal inhibitory concentration (MIC) breakpoints for staphylococcus spp[S].National Committee for Clinical Laboratory Standards.M100-S14.2004:24(1):42.
  • 5MIC interpretive standards(μg/ml) for staphylococcus spp[S].National Committee for Clinical Laboratory Standards.M100-S14.2004:24(1):105.
  • 6Louie L,Matsumura SO,Choi E,et al.Evaluation of three rapid methods for detection of methicillin resistance in staphylococcus aureus[J].J Clin Microbiol,2000,38(6):2170-2173.
  • 7Mackenzie AM,Richardson H,Lannigan R,et al.Evidence that the National Committee for Clinical Laboratory Standards disk test is less sensitive than the screen plate for detection of low-expression-class methicillin-resistant Staphylococcus aureus[J].J Clin Microbiol,1995,33(7):1909-1911.
  • 8Rohrer S,Maki H,Berger-Bachi B.What makes resistance to methicillin heterogeneous[J].J Med Microbiol,2003,52(8):605-607.
  • 9张庆玲,刘明华,王仙园.耐甲氧西林金黄色葡萄球菌的研究进展[J].中华医院感染学杂志,2002,12(10):797-798. 被引量:51

共引文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部