医用^(60)Co治疗机吸收剂量测量不确定度分析 被引量:4

Measurement uncertainty evaluation of absorbed dose for ^(60)Co teletherapy equipment
下载PDF
导出
摘要 针对原IAEA 1987年版和IAEA 1986年版TRS.277中吸收剂量测量的不确定度评定与实际相差过大,不能满足临床应用时5.0%(k=2)要求的状况,依据IAEA 1996年第二版和1998年第二版的TRS.277号报告,参考TRS.398、TRS.381及其他相关技术资料,对医用60Co远距离治疗辐射源吸收剂量测量不确定度进行新的分析及评定,得到约为3.0(%k=2)的吸收剂量测量不确定度。实验结果表明,分析合理,其结果符合实际情况,满足临床应用要求。 In the IAEA TRS.277 version 1987 and 1986, the absorbed dose measurement uncertainty evaluation is of great difference from the actually situation, and can not meet the clinical requirement of 5.0% (k=2). The measurement uncertainty of absorbed dose for ^60Co teletherapy equipment was re-analyzed and re-evaluated according to the TRS.277 2nd edition(both 1996 2nd edition and 1996 2nd edition), referring to the TRS.398 and TRS.381 and other correlative data. The obtained uncertainty of dose measurement was about 3.0%(k=2). The experimential results indicate that the analysis and evaluation in our work is reasonable, and the uncertainty complies with the actual situation, and matches the clinical requirement.
出处 《中国测试技术》 2008年第4期39-43,共5页 CHINA MEASUREMENT & TESTING TECHNOLOGY
关键词 TRS.277 ^60Co远距离治疗机 有效测量点 吸收剂量 测量不确定度 TRS.277 ^60Co teletherapy equipment Effective measuring point Absorbed dose Measurement uncertainty
  • 相关文献

参考文献11

  • 1IAEA. Absorbed dose determination in photon and electron beams:an international code o.f practice (2nd edition, tech. Per.ser.No.277) [S]. Vienna: IAEA, 1996.
  • 2IAEA. Absorbed dose determination in photon and electron beams: an international cede of practice (2nd Edition, Tech.Per.ser.no.277) [S]. Vienna: IAEA, 1998.
  • 3JJF1059-1999,测量不确定度评定与表示[S].北京:中国计量出版社,1999.
  • 4JJG590-1989,^60Co远距离治疗辐射源[S].北京:中国计量出版社,1989.
  • 5JJG589-2001,外照射治疗辐射源[S].北京:中国计量出版社,2001.
  • 6IAEA. The use of plane-parallel ionization chambers in high -energy electron and photon beams:an international code of practice for dosimetry (Tech. Per. ser. no. 381)[S]. Vienna: IAEA,1995.
  • 7IAEA. Absorbed dose determination in external beam radiotherapy:an international code of practice for dosimetry base on standards of absorbed dose to water (Tech. Per. ser. No. 398) [S]. Vienna: IAEA,2001.
  • 8JJG912-1996,治疗水平电离室剂量计[S].北京:中国计量出版社,1996.
  • 9黄扬.治疗水平电离室剂量计校准因子检定测量不确定度评定[J].中国计量,2006(12):67-69. 被引量:4
  • 10ICRU. Use of computers in external beam radiotherapy procedures with high-energy photons and electrons (ICRU REPORT 42) [S]. Baltimore: ICRU,1987.

共引文献5

同被引文献30

  • 1黄扬,杨勇,张圆月,鄢铃,魏川.X、γ射线照射量(治疗水平)标准装置技术研究[J].中国测试技术,2005,31(6):32-35. 被引量:6
  • 2李大鹏,余辉,闫微,郑峰,沈增贵.后装治疗机的现状及其在宫颈癌中的应用[J].医疗设备信息,2007,22(3):65-67. 被引量:17
  • 3ICRU Report 24:1976,Determination of absorbed dose in a patient irradiated by beams of X or gamma rays in radiotherapy procedures[S].International Commission on Radiation Units and Measurements Bethesda,ICRU,1976.
  • 4IAEA Technical Reports Series No 277:1996,Absorbed dose determination in photon and electron beams:an international code of practice[S].2nd Edition.Vienna,IAEA,1996.
  • 5IAEA Technical Reports Series No 277:1998,Absorbed dose determination in photon and electron beams:an adaptation of the IAEA international codes of practice[S].2nd Edition.Vienna,IAEA,1998.
  • 6IAEA Technical Report Series No 381:1997,The use of plane-parallel ionization chambers in high-energy electron and photon beams.An International Cede of Practice for Dosimetry[S].Vienna,IAEA,1997.
  • 7IAEA Technical Report Series No 398:2000,Absorbed dose determination in external beam radiotherapy.An International Cede of Practice for Dosimetry Based on Standards of Absorbed Dose to Water[S].Vienna,IAEA,2000.
  • 8IAEA Technical Report Series No 398:2004,An International code of preactice for Dosimetry based on absorbed dose to water,Absorbed dose determination in external beam radiotherapy[S].Vienna,IAEA,2004.
  • 9JJG912-1996,治疗水平电离室剂量计[S].北京:中国计量出版社,1997.
  • 10JJG 589-2001,外照射治疗辐射源[S].北京:中国计量出版社,2001.

引证文献4

二级引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部