16层螺旋CT在诊断冠状动脉病变中应用价值

The Value of Sixteen-slice Spiral CT Angiography of Coronary Artery in the Diagnosis of Coronary Heart Disease
下载PDF
导出
摘要 目的探讨16排螺旋CT在冠状动脉病变诊断中的价值。方法和材料对60例临床怀疑冠心病的患者进行心脏冠状动脉CT成像。男32例,女28例,年龄28 ̄76岁,平均54.8岁。所有患者进行了传统冠状动脉造影术做对照,分析16排螺旋CT冠状动脉成像检查的灵敏度、特异度。结果16排螺旋CT冠状动脉成像检查与冠状动脉造影相比,灵敏度为81.35%,特异度为96.43%。结论16层螺旋CT冠状动脉成像技术是一种无创、简便的检查方法,显示有临床意义的冠状动脉狭窄的准确性很高,作为一项非创伤性技术,是评价冠状动脉狭窄病变的重要筛选手段。 Objective To evaluate the value in diagnosing coronary disease with 16 spiral CT. Methods and Materials A total of 60 suspected coronary artery disease patients were checked by 16-slice CT,including male 32 cases,female 28 cases,and the average age was 54.8 years old.All of them also were checked by vonventional coronary angiogTaphy to analyze the sensitivity and specificity of 16-slice spiral CT coronary artery imaging. Results Among the 60 cases,the sensitivity was 81.35% and the specificity was 96.43%. Conclusion 16-slice CT allows the accurate,non-invasive detection of significant coronary stenosis.h provides a good visualization. It is valuable in demonstrating coronary lumen stenosis.It should be used as a reliable exam of coronary heart disease in clinial practice.
出处 《中国医疗设备》 2008年第7期131-133,共3页 China Medical Devices
关键词 体层摄影术 冠状血管造影术 冠状动脉疾病 螺旋CT tomography coronary angiography coronary disease sprial CT
  • 相关文献

参考文献3

二级参考文献15

  • 1De Feyter PJ,Nieman K,van Ooijen P,et al.Non-invasive coronary arteryimaging with electron beam computed tomography and magnetic resonance imaging[J].Heart,2000,84(4):442
  • 2Moshage WE,Achenbach S,Seese B,et a1.Coronary artery stenosis:three-dimensional imaging with electrocardiographically triggered,contrast agent-enhanced electron-beam CT[J].Radiology,1995,196(4):707
  • 3Rubin GD,Dake MD,Semba CP.Current status of three-dimensional spiral CT scanning for imaging the vasculature.Radiol Clin North Am,1995,33:51-70.
  • 4Rubin GD,Silverman SG.Helical (spiral) CT of the retroperitoneum.Radiol Clin North Am,1995,33:903-932.
  • 5Becker CR,Knez A,Ohnesorge B,et al.Imaging of noncalcified coronary plaques using helical CT with retrospective ECG gating.AJR,2000,175:423-424.
  • 6Hong C, Becker CR, Huber A,et al.ECG-gated reconstructed multi-detector row CT coronary angiography:effect of varying trigger delay on image quality.Radiology,2001,220:712-717.
  • 7Nieman K, Cademartiri F, Lemos PA, et al. Reliable noninvasive coronary angiography with fast submillimeter multislice spiral computed tomography E J ]. Circulation, 2002,106(16) : 2 051-2 054
  • 8Wang G, Vanner MW. The effect of pitch in multislicespiral/helicla CT[J]. Med Phys, 1999, 26:2 648-2 653
  • 9Achenbach S, Giesler T, Ropers D, et al. Detection of Coronary Artery Stenoses by Contrst - Enhanced, Retrospectively Electrocardiographically - Gated, Multislice Spiral Computed Tomography [ J ]. Circulation, 2001, 103(21) : 2 535-2 538
  • 10Flohr TH, Stierstorfer K, Bruder H, et al. Image reconstruction and image quality evaluation for a 16 - slice CT scanner [ J ]. Medical Physics, 2003,30 (5) : 832-845