Sysmex KX-21N半自动血细胞分析仪的使用及评价

Evaluation and employment of sysmex KX-21N semi-automatic blood cell analyzer
下载PDF
导出
摘要 目的:对sysmex KX-21N半自动血细胞分析仪进行使用评价。方法:采用随机选择方法对该机器进行批内精密度、相关性、对比性进行了测试。结果:仪器精密度试验,批内CV值在0.61%~3.5%之间,用全血法和预稀释法试验相关性比较,各指标相关系数在0.9406~0.981之间。结论:该仪器主要指标评价结果符合设计性能,适用于我院检验科常规血液分析。 Objective:To evaluate the function of sysmex KX-21N semi-automatic blood cell analyzer, method:assay the within-batch precision, correlation, relative property of the instrument with the random select method. Result:The CV is between 0.61%-5.5% in degree of precision test of equipment; To compare with whole blood method and force-dilution method, The coefficient correlation of each index is between 0.9406%-0.981%. conclusion:The main index of the instrument is consistent with the design and fit for routine blood analysis in our hospital.
机构地区 黑龙江省医院
出处 《中国医学装备》 2008年第8期14-15,共2页 China Medical Equipment
关键词 相关性 批内精密度 correlation, within-batch precision