摘要 目的评估XE-2100全自动血液分析仪对异常白细胞报警的可信度。方法对105例XE-2100全自动血液分析仪有异常白细胞提示的标本及200例无异常细胞报警的标本进行手工推片、瑞氏染色镜检。结果XE-2100全自动血液分析仪对异常细胞报警的灵敏度为100%,特异性为85.1%,阳性预测值为66.7%,阴性预测值为100%。结论XE-2100全自动血液分析仪作为异常细胞的初筛工具,其灵敏度极高,仍不能完全代替手工镜检。 Objective To evaluate the reliability of white blood cell morphologic abnormalities warnings of system XE -2100 hematology analyzer. Method Manual stained differentiation was performed on 200 normal morphology samples and 105 abnormal morphology samples. Resuits The analyzer test showed that the sensitivity and the specificity were 100 % and 85.1% for abnormal morphology, with the positive predictive value of 66. 7 % and the negative predictive value of 100 %. Conclusions The system XE - 2100 hematology analyzer has good primary screening ability and higher sensitivity, but it cannot completely take place of manual microscopic examination.
出处 《实用预防医学》 CAS 2008年第4期1239-1240,共2页 Practical Preventive Medicine
关键词 血液分析仪 异常白细胞 报警 Hematology analyzer Abnormal white blood cell Warning
  • 相关文献

参考文献10

二级参考文献16

  • 1冯仁丰,黄维纲,王晓明,金大鸣,徐丽安,倪赞明.K-4500血液分析仪使用探讨[J].中华医学检验杂志,1996,19(2):112-114. 被引量:25
  • 2Darnige L Cluet-Dennetière S Delavenne J et al.Evaluation de I''analyseur d''hématologie Beckman Coulter HmXTM dans un hopital général[J].Annales de Biologie Clinique,2002,60(1):44-55.
  • 3R Herklotz, AR Huber. Precision and accuracy of the leukocyte differential on the Sysmex XE-2100 [J ]. Sysmex J Int,2001,11:8-21.
  • 4H Inoue. Overview of automated hematology analyzer XE-2100[J]. Sysmex J Int,1999,9:58-64.
  • 5Fujimoto K. Principles of measurement in hematology analyzers manufactured by sysmex corporation [J]. Sysmex J Int, 1999, 9:31- 40.
  • 6Hohenwallner W, Wiesinger K, Wimmer E. The reticulocytes: automatic counting, indication and interpretation[J]. Sysmex J Int, 1992, 2:120-135.
  • 7Carol B, Paul H, Donna G. Performantion evaluation of the sysmex XE-2100 automated hematology analyzer [J]. Sysmex J Int, 1999,9:113-119.
  • 8Jerelyn W, Patracia G. Performance evaluation of sysmex XE-2100 hematology analyzer[J]. Lab Hematol,2000, 6:83-92.
  • 9National Committee for Clinical Laboratory Standards.Reference leukocyte differential count (proportional) and evaluation of instrument methods: approved standard[Z]. NCCLS Document H20-A. NCCLS, Wayne (USA), 1992.
  • 10Sysmex公司仪器的技术指标.Sysmex XE-2100中文操作手册[Z],1999,1:14.

共引文献19

同被引文献5

引证文献1

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部