摘要
目的:观察薄层扫描法中不同峰宽和斜率值对扫描结果的影响。方法:依次改变峰宽和斜率值对斑点进行扫描,比较含量测定结果。结果:峰宽和斜率值对含量测定影响较大,最佳峰宽-斜率组合为(1.30,0.75),此时方程r值为1.000,测得值为理论值的99.27%。结论:薄层扫描法参数中应注明峰宽和斜率值。
Objective:To observe the effect on determination of peak width and slope in TLC scanning. Methods:To scan the dot after the value of peak width and slope changed in order, and compare the determined content. Results:The content was much effected by peak width and slope, the best couple of them were (1.30, 0.75), its formula's value of r was 1.000, the content was equal to 99.27% of theoretical content. Conclusion:The value of peak width and slope ought to be marked during the index of TLC scanning.
出处
《中国医药导报》
CAS
2008年第23期127-127,130,共2页
China Medical Herald
关键词
薄层扫描法
峰宽
斜率
TLC scanning
Peak width
Slope