期刊文献+

OrbscanⅡ角膜地形图仪和A型超声角膜测厚仪测量LASIK手术前后角膜厚度的比较 被引量:5

Measurements of corneal thickness with the OrbscanⅡ Topography System and with A-ultrasonic pachymeter
下载PDF
导出
摘要 目的比较OrbscanⅡ角膜地形图仪和角膜测厚仪(A超)检测角膜厚度的差异。方法分别用Orb-scanⅡ角膜地形图及A超对角膜厚度进行检测,并将测量值进行统计学分析。结果OrbscanⅡ测得术前角膜厚度为(524.97±41.41)μm,A超测得角膜厚度为(528.57±33.04)μm。OrbscanⅡ测量值与A超测量值差异有统计学意义(t=-5.093,P<0.001),两种测厚结果高度正相关(r=0.915,P<0.001),回归方程如下:A超测量值(μm)=158.840+[0.702×OrbscanⅡ测量值(μm)]。通过检验,两种测厚方法对术后角膜厚度的测量结果符合术前的回归方程。结论OrbscanⅡ和A超检测中央角膜厚度显示了高度的相关性;二者结合使用可以为LASIK手术患者提供准确的角膜厚度值,并可从总体上了解角膜的形态,可以准确地评估屈光手术前后角膜的参数,并有效保证手术的安全性。 Objective To compare the difference measurements of corneal thickness with the Orbsean Ⅱ Topography system and with A-ultrasonic pachymeter. Methods To measure 967 eyes of 490 myopia patients with the Orbscan Ⅱ Topography System and A-ultrasonics respectively. Then compare the result of corneal thickness. Results The mean center corneal thickness was (524.97 ± 41.41 )μm with the Orbscan Ⅱ System and (528.57 ± 33.04 )μm with the A-ultrasonics,these values being signifieandy diferent( t = -5. 093, P 〈 0.01 ). There were significant linear correlations between Orbscan Ⅱ and A-ultrasonic( r =0.915, P 〈0.001 ). The regression equation was :Y( A-ultra sonic, μm) = 158. 840 + [. 702 x X ( Orbscan Ⅱ, μm ) ]. Conclusion There was a significant correlation between the results otained by both techniques. Contrary to the ultrasonic pachymeters the Orbsean simultaneously measures several thousand points including the thinnest point and provides a colour-coded thickness map without physical contact.
出处 《临床眼科杂志》 2008年第4期302-304,共3页 Journal of Clinical Ophthalmology
关键词 角膜厚度 近视眼 ORBSCAN Ⅱ角膜地形图 A型超声 Crneal thickness Mopia A-ultrasonic Obscan Ⅱtopography system
  • 相关文献

参考文献9

  • 1Joo CK, Kim TG. Corneal ectasia detected after laser in situ keratomileusis for correcton of less than 12 diopters of myopia. J Cataract Refract Surg, 2000,26:292-295.
  • 2Seller T, Koufala K, Richter G. latrogenic keratectasia after laser in situ keratomileusis. J Refract Surg, 1998,14:312-317.
  • 3McLeod SD, Kisla TA, Caro NC, et al. latrogenlc keratoconus : corneal ectasia following laser in situ keratomileusis for myopia. Arch Ophthalmol 2000,118:282-284.
  • 4Yaylali V, Kaufman SC,Thompson HW. Corneal thickness measurements with the Orbscan Topography System and Ultrasonic pachymetry. J Cataract Retract Surg, 1997,23 : 1345-1350.
  • 5Iskander NG,Anderson Penno E,Peters NT,et al. Accuracy of Obscan pachymetry measurements and DHG ultrasound pachymetry in primary laser in situ keratomilensis and LASIK enhancement procedures. J Cataract Refract Surg,2001,27 :681-685.
  • 6Rainer G, Findl O, Petternel V, et al. Central corneal thickness measurements with partial coherence interferometry, ultrasound, and the Orbsean system. Ophthalmology ,2004,111:875-879.
  • 7Chakrabarti HS, Craig jP, Brahma A, et al. Comparison of corneal thickness measurements using ultrasound and Orbscan slit-scanning topography in normal and post-LASIK eyes. J Cataract Refract Surg,2001,27 :1823-1828.
  • 8Wong AC, Wong CC, Yuen NS, et al. Correlational study of central corneal thickness measurements on Hong Kong Chinese using optical coherence tomography, Orbscan and ultrasound pachymetry. Eye, 2002,16:715-721.
  • 9Kawana K, Hiraoka T, Okamoto F, et al. Comparison of 3 different corneal thickness measurements in eyes with keratoeonus. Invest Ophthalmol Vis Sei ,2004,45:2892.