摘要
目的:应用全自动生化分析仪建立缺血修饰白蛋白(IMA)的生物学参考范围。方法:在Hitachi7170A全自动生化分析仪上设计合理参数,采用白蛋白钴结合(ACB)试验测定IMA,测定561例健康人群血清IMA含量,建立健康人群IMA参考范围。结果:健康人群中小于30岁人群的5%参考下限为67 U/ml,>30岁人群的5%参考下限为64 U/ml。结论:用全自动生化分析仪建立ACB方法测定的IMA参考范围,有助于临床对心肌缺血的诊断。
Objective:To establishe the reference intervals of IMA on automatic biochemical analyzer. Methods: IMA were detected by Hitachi 7170A autobiochemical analyzer with albumin cobalt binding test. IMA reference intervals were established by 561 healthy controls. Results:5% lower limit values in the subjects lower and old 30 years were 67 U/ml and 64 U/ml respectively. Conclusions:The reference intervals of IMA are established on automatic biochemical analyzer with ACB test. It should be helpful for clincal diagnosis on myocardial ischemia.
出处
《中国误诊学杂志》
CAS
2008年第26期6301-6303,共3页
Chinese Journal of Misdiagnostics
关键词
血清白蛋白/分析
北京
汉族
参考值
人类
Serum Albumin/analysis
BEIJING
HAN NATIONALITY
Reference Values
Humans