期刊文献+

多普勒、心导纳与心导管法测定小儿室间隔缺损心输出量的比较

PULSED DOPPLER, DIFFERENTIAL-ADIMITTANCE CARDIOGRAM AND FICK METHOD DETERMINATION OF CARDIAC OUTRUT IN CHILDREN WITH VENTRICULAR SEPTAL DEFECT
下载PDF
导出
摘要 本文应用脉冲多普勒超声心动图(简称多普勒)和心导纳微分图微电脑自动分析(简称心导纳)测定27例小儿单纯性空间隔缺损心输出量(CO)和心脏指数(CI),与心导管(Fick)法相比,多普勒法CO和CI相关系数分别为0.92,0.90;心导纳法CO和CI相关系数分别为0.38、0.24,表明多普勒能准确测定心导纳不宜测定的小儿室间隔缺损CO和Cl。 The validity of pulsed Doppler and differentila-admittance cardiogram of cardiac output (CO) and cardiac index(CI) of ventricular septal defect, was compared with Fick's method in 27 children of 3-14 years old. CO and CI of pulsed Doppler correlated well with that of Fick(r=0. 92,0. 90) ; but differential-admittance cardiogram not well correlated(r = 0. 38,0. 24). This Study indicated that CO and CI of children with ventricular septal defect can be determinated by pulsed Doppler as a noninva-sive approach accurately but the differential admittance cerived ones is unacceptable.
出处 《湖南医科大学学报》 CSCD 1990年第4期366-368,共3页 Bulletin of Hunan Medical University
关键词 脉冲多普勒 室间隔缺损 心脏指数 pulsed Doppper differential-admittance cardiogram heart septal defects ven- tricle cardiac output cardiac index
  • 相关文献

参考文献1

二级参考文献3

  • 1谢振武,李茗香,郭青,李冠平,谢林军.健康儿童心阻抗微分图及其相关因素分析[J]湖南医学,1988(04).
  • 2A. R. S. Bukhari,K. Yamakoshi,T. Toyoshima,T. Togawa,H. Ito. Three-channel telemetry system for monitoring e.c.g. cardiac output, rate of respiration and basal transthoracic admittance[J] 1981,Medical & Biological Engineering & Computing(4):426~432
  • 3D. W. Hill,F. D. Thompson. The importance of blood resistivity in the measurement of cardiac output by the thoracic impedance method[J] 1975,Medical & Biological Engineering(2):187~191