彩色多普勒血流系统中血流参数的算法分析

The Arithmetic Analysis of Flow Parameter in the Color Flow Mapping System
下载PDF
导出
摘要 目的本文通过讨论处理多普勒信号的常用算法:自相关法,AR模型法,希望找到一种性能较好的算法。方法利用仿真数据以及取自人体的真实数据对算法进行仿真比较。结果通过几组仿真数据的比较可看出,二者的中心频率估计差别不大,但AR模型法的分辨率及方差性能明显优于自相关法。由于CFM计算时数据较短(通常为8~16个),因此用自相关法估计方差时,谱展宽现象严重。结论经典功率谱估计方法的方差性能较差,分辨率较低。 Objective To explore a better arithmetic of Doppler signal by discussing lag correlation method, AR model method. Methods Compare the two methods by simulation data and truthful data of human body. Results Center frequency of two methods doesn't have obvious difference, but resolving capability and skedasticity of AR model method was better than lag correlation methods. Because the data was shorter calculated by CFM (usual 8-16), spectrum widening was badly estimated by lag correlation method. Conclusion Scedasticity of classic power spectrum estimation was bad,resolving capability was bad.
出处 《中国现代医生》 2008年第24期19-21,共3页 China Modern Doctor
关键词 多普勒信号 自相关法 AR模型法 Doppler signal Lag correlation method AR model method
  • 相关文献

参考文献5

  • 1[1]Schlindwein FS,Evans DH.A real-time autoregressive spectrum analyzer for Doppler ultrasound signals[J].Ultrasound Med Biol,1989,15(3):263-272.
  • 2[2]Hein IA,O'Brie WR.Current time-domain methods for assessing tissue motion by analysis from reflected ultrasound echoes-a review[J].IEEE Trans Ultrason Ferroelectr Freq Control,1993,40(2):84-102.
  • 3[3]Jensen JA.Stationary echo canceling in velocity estimatlon by time-domain cross-correlation[J].IEEE Trans Med Imaging,1993,12(3):471-477.
  • 4[4]Jensen JA.Implementation of ultrasound time-domain cross-correlation blood velocity estimators[J].IEEE Trans Biomed Eng,1993,40(5):468-474.
  • 5[5]Keeton PI,Schlindwein FS.Spectral broadening of clinical Doppler signals using FFT and autoregressive modelling[J].Eur J Ultrasound,1998,7(3):209-218.

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部