摘要
采用流动注射技术,研究了延胡索乙素-高锰酸钾-连二亚硫酸钠体系的化学发光行为,对影响化学发光强度的诸因素进行了实验和探讨,建立了化学发光法测定延胡索乙素的新方法。方法的检出限为3.2 ×10-9g/mL,线性范围为1×10-8~8×10-6g/mL。对罗通定中延胡索乙素含量的测定结果满意,并与药典标准方法测得值一致。
A new method for the determination of tetrahydropalmatine using flow injection chemiluminescence was studied. It is based on the chemiluminscence reaction of tetrahydropalmatine-KMnO4-Na2S2O4. The main factors which affect the determination were investigated. Under optimum conditions, the linear range for the determination of tetrahydropalmatine was 1 × 10-8~8 × 10-6 g/mL and the determination limit was 3. 2 × 10-9 g/mL. The method has been applied to the determination of tetrahydropalmatine in Rotundine. tablets and their reults are in good agreement with those obtained by classical method.
出处
《分析化学》
SCIE
EI
CAS
CSCD
北大核心
1997年第11期1321-1323,共3页
Chinese Journal of Analytical Chemistry
基金
国家自然科学基金
关键词
延胡索乙素
化学发光
流动注射
罗通定
元胡
Tetrahydropalmatine, chemiluminscence, flow injection technique