摘要
在酸性条件下,Ce(Ⅳ)氧化Ru(bipy)3^2+生成Ru(bipy)3^3+,同时氧化氧氟沙星生成中间态的自由基,Ru(bipy)3^3+和中间态自由基接触后能产生强烈的化学发光.利用该反应,将Ru(bipy)3^2+固定在阳离子交换树脂上,结合流动注射分析技术,建立了一种测定氧氟沙星的流动注射化学发光传感器.结果表明,氧氟沙星浓度在2×10^-7~8×10^-6mol/L(r=0.999 3)范围内与化学发光强度有良好的线性关系,检出限为6×10^-8mol/L(3σ),对5×10^-6mol/L的氧氟沙星测定的相对标准偏差(RSD)为2.2%(n=11).该传感器成功地用于血清样品中的氧氟沙星的测定.
A novel flow-injection chemiluminescence sensor for determination of ofloxacin (OFLX) is described in this paper. It was prepared by immobilizing the Ru(bipy)2^2+ on a cation exchange resin column. OFLX and Ru(bipy)3^2+ were oxided both by Ce(N) into Ru(bipy)3^2+ and intermediate radical in acidic condition. Then, OFLX was sensed by the chemiluminescence(CL) reaction based on the reaction between Ru(bipy)3^2+ and intermediate radical. The linear range of OFLX concentration is 2 × 10^-7~8 × 10^-6 mol/L(r=0.999 3). The detection limit was 6 × 10^-8 mol/L (3δ) and the RSD was 2.2M (5 × 10^-6mol/L,n=11). This sensor has been successfully applied to determine ofloxacin in serum.
出处
《陕西师范大学学报(自然科学版)》
CAS
CSCD
北大核心
2008年第5期48-51,共4页
Journal of Shaanxi Normal University:Natural Science Edition
基金
国家自然科学基金资助项目(30470886)