医用直线加速器治疗床对放疗剂量影响的研究 被引量:7

Research of Effect for Medical Linear Accelerator Treatment Table on Radiotherapy Dosimetry
下载PDF
导出
摘要 目的探讨西门子医用直线加速器治疗床对放疗剂量影响的研究。方法将固体水模固定在治疗床中心处,改变加速器机臂的角度,从不同角度照射并用剂量仪进行比对测量,计算出治疗床不同床板对放疗剂量的衰减因子。结果西门子医用直线加速器治疗床中有机玻璃床头板对剂量的衰减在1.5~20.5%,有机玻璃床体板对剂量的衰减在1.9~38.6%,有机玻璃网状窗体板对剂量的衰减在0.18~12.3%。结论放射治疗时正后野180°左右时应选用网状板或有机玻璃床头板,在后侧斜野110°~130°和230°~250°区间最好用床尾板,并对相应剂量做出修正与补偿,而治疗床的主支撑架及金属边柜对剂量的衰减高达12.3~38.6%,而床头延伸板对剂量的衰减在1.1~5.5%之间,因此头部的肿瘤应选此延伸板治疗。因此在治疗计划设计与实施过程中应避免射线束直接穿过主支撑架与金属边框。 Objective To investigate the effect of radiotherapy dosage caused by the SIEMENS medical linear accelerator table. Methods The solid water phantom was fixed in the center of the table , and changed the angle of the machine arm. The chamber dosimeter was used to measure the dose from different angle , the factor of dosage declined by the different table planks was calculated. Result The radiotherapy dose was declined on the average by 1.5 -20.5 %, 1.9 -38.6%, and 0.18 - 12.3 % when the plexiglass table head plank , plexiglass table body plank and plexiglass web winder body plank in the SIEMENS medical linear accelerator table were used re- spectively. Conclusions The web table plank or plexiglass table head plank must be chose in the radiotherapy of the positive back field about 180°, in the back -flank field about 110° - 130° or 230° -250°chose the table tail plank is the best method , and should be rectified and revised by dose compensate . When beam passed through the metal handle and main support shelf of the table, it would reduce the dose by 12.3% -38.6% ,but the head ex- tending table plank reduce the dose by - 1.1% - 5.5 %, so it must be chose in the head tumor radiotherapy. In the process of designing and practicing the treatment plan should keep the beam away the metal handle of the bed or the mental support of the table plank .
出处 《生物医学工程学进展》 CAS 2008年第3期161-163,共3页 Progress in Biomedical Engineering
关键词 加速器 治疗床 放疗 剂量 影响 Accelerator Treatment table Radiotherapy Dosimetry Effect
  • 相关文献

参考文献2

  • 1[3]Klein EE,Chin LM,Rice RK,et al.The infuluence of the air cavities oninterface doses for photon beams[J].Int J Radiat Oncol Biol Phys,1993,27:419-427
  • 2[4]Mian TA,van Putten MC,Kramer DC,et al.Backscatter ra-diation at bone-titanium intedace from high-energy Xand gamma rays[J],lnt J Radiat Oncol Biol Phys,1987,13:1943-1947

同被引文献28

  • 1盛延兴,隋振忠,郑芳霞,李学章,魏绪国,于书增.立体定向放疗床、头架、真空袋对吸收剂量的影响[J].现代肿瘤医学,2004,12(3):184-184. 被引量:16
  • 2李兆斌,徐利明,王晓红,李承军.真空定位袋的质量保证与质量控制[J].中国医学物理学杂志,2006,23(5):317-319. 被引量:15
  • 3Klein EE,Chin LM,Rice RK,et al. The infuluence of the air cavities on interface doses for photon beams [J]. Int J Radiat Oncol Biol Phys, 1993,27(2) :419-427.
  • 4Mian T A, Van Putten M C J r, K ramer D C, et al. Backscatter Radiation at Bone-titanium Interface from High-energy X and Gamma Rays[J]. Int J Radiat Oncol Biol Phys,1987,13(12): 1943-1947.
  • 5胡逸民,张红志,戴建荣.物理技术方面质量保证.肿瘤放射物理学[M].北京:原子能出版社,1999.618-636.
  • 6胡逸民,张红志,戴建荣.放射治疗的质量保证与质量控制.肿瘤放射物理学.北京:原子能出版社,1999:612-626.
  • 7Christine SP, Ghyslain L, Charles L, et al. Correction of beam attenuation caused by carbon fiber tabletop in treatment planning system for radiotherapy[J]. Med Phys, 2009, 25(5): 547-552.
  • 8Poppe B, Chofor N, Ruhmann A, et al. The effect of a carbon-fiber couch on the depth-dose curves and transmission properties for megavoltage photon beams[J]. Strahlenther Onkol, 2007, 183(1): 43 -48.
  • 9Myint WK, Niedbala M, Wilkins D, et al. Investigating treatment dose error due to beam attenuation by carbon fiber tabletop[J]. Med Phys, 2006, 7(3): 21-27.
  • 10McCormack S, Diffey J, Morgan A, et al. The effect of gantry angle on megavoltage photon beam attenuation by a carbon fiber couch insert[J]. Med Phys, 2005, 32(2): 483-487.

引证文献7

二级引证文献47

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部