期刊文献+

基于高频X线机的DR升级系统的应用 被引量:8

The Application of DR Upgrade System Based on X-ray Equipment of High Frequency
下载PDF
导出
摘要 目的探讨应用平板探测器升级高频X光机成为DR系统的方法及其应用价值。方法利用CANON公司的CXDI-50G平板探测器升级PHILIPS Bucky Ts高频X光机,采集的图像信息传送工作站进行图像处理。结果图像质量满足诊断要求,与医院PACS、RIS系统联网成功。结论基于高频X光机的DR升级系统,提高了影像质量,降低了材料损耗和购机成本,有一定的推广和应用价值。 Objective To evaluate the method and application of DR upgraded system based on high fre- quency X - ray equipment by using fiat panel detector. Methods To upgrade PHILIPS Bucky Ts X - ray equip- ment of high frequency by using CXDI -50G fiat panel detector of CANON company,The acquired image informa- tion was transmitted to the post processing workstation. Results The image qulity was satisfied for the diagnosis request. DR upgraded system was successful connected between the PACS and R_IS system. Conclusion DR up- graded system improved image quality , reduced the cost of purchase and had extensive significance in the applica- tion.
出处 《生物医学工程学进展》 CAS 2008年第3期179-180,共2页 Progress in Biomedical Engineering
关键词 平板探测器 X光机 DR 升级 Flat panel detector X - ray equipment DR Upgrade