摘要
采用指数修正的高斯模型(EMG),研究了进样时间、分离电压及柱长等一系列操作条件对毛细管区带电泳峰形参数的影响。实验结果表明,进样时间对半峰宽(W1/2)和峰不对称度(B/A)有显著影响;电压对以长度表示的半峰宽及峰拖尾参数无显著影响;tm与σ、τ、W1/2(s)之间都有很好的线性关系;随着柱长的增大,σ、τ、W1/2都略有增加。
The influence of operation conditions, such as, injection time, applied voltage and effective length of capillary column, on parameters of peak shape in capillary zone electrophoresis (CZE) was investigated by using exponential modified Gaussian (EMG) equation. The CE experiments were carried out by using the sample of derivatives of benzoic acid and small peptides. The experimens indicated that there was a significant effect of injection time on W 1/2 and that the peaks changed from tailing to leading when the injection time was over 3 seconds in the case of pressure injection. It is recommended that an injection time less than 3 seconds can lead to a good column efficiency. The good linear relationships between t m and τ, σ,W 1/2 (t) with different separation voltages were observed. In the range of experimental conditions (11 27kV), higher voltage could lead to smaller zone dispersion, giving rapid and high efficient separation, and the voltage was insignificant on peak tailing. It was also shown that σ, τ and W 1/2