摘要
用冰冷却水吸收二甲醚配制样品,将TCD和FID串联使用,测定出二甲醚在TCD和FID上相对于甲醇的重量校正因子分别为0.86和0.55,并通过用甲醇催化脱水生成的二甲醚和水的化学计量关系,在线间接测定二甲醚在TCD上相对于甲醇的重量校正因子,从而使上述测定结果得到佐证。
Dimethyl ether(DME) correction factors in gas chromatography with thermal conductivity detector(TCD) and flame ionization detector(FID) by using H 2 as carrier gas were determined in this work. The home made DME gas was quantitatively absorbed in ice cold water. With ethanol as standard, the aqueous mixture was injected into a gas chromatograph, equipped with serially connected TCD and FID. The weight correction factors of DME based on methanol were 0 86 and 0 55 for TCD and FID respectively. The result for TCD was also confirmed by calculation based on the stoichiometrical transformation of methanol into DME in reaction gas chromatography.
出处
《色谱》
CAS
CSCD
北大核心
1997年第3期248-249,共2页
Chinese Journal of Chromatography