期刊文献+

流动注射化学发光法测定杀扑磷 被引量:2

The Flow-injection Chemiluminescent Analysis for the Determination of Methidathion
下载PDF
导出
摘要 基于杀扑磷对过氧化氢和鲁米诺在碱性条件下化学发光有抑制作用,本文建立了一种杀扑磷的流动注射化学发光(FI-CL)检测方法。杀扑磷在1.0×10-7~1.0×10-3g/mL范围内与发光强度呈良好的线性关系,检测限(S/N=3)为7.3×10-8g/mL,测定相对标准偏差(RSD)为1.26%,对柑橘样品进行加标回收率测定,回收率在89.5%~92.6%之间。 A method for the determination of methidathion based on the inhibition of chemilumineseenee(CL) of luminol-H2O2 in basic medium by flow-injection CL analysis has been established. The calibration curve was linear for the methidathion range of 1.0× 10^-7-1.0 × 10^-5g/mL with a detection limit(S/N=3) of 7.3 ×10^-8g/mL and the relative standard (RSD) was 1.26%. The method was applied to the determination of methidathion in orange and the recoveries were between 89.5% -92.6%.
出处 《福建分析测试》 CAS 2008年第4期10-13,共4页 Fujian Analysis & Testing
关键词 流动注射 化学发光 杀扑磷 flow-injection chemiluminescence methidathion
  • 相关文献

参考文献4

二级参考文献16

  • 1应兴华,徐霞,陈能,朱智伟.气相色谱法快速测定柑桔汁中残留杀扑磷[J].食品工业科技,2006,27(4):166-167. 被引量:2
  • 2孙彩霞,姚晗珺,赵志强,刘璇.浙江省柑橘应对日本“肯定列表制度”研究[J].浙江农业科学,2006,47(6):605-608. 被引量:5
  • 3GB15670-1995.农药登记毒理学试验方法[S].[S].国家技术监督局,..
  • 4P.Martos,J.Pawliszyn,Anal.Chem.,1999,71:1513.
  • 5Hanwen Wang,Wenmin Liu,Yafeng Guan.LCGC North America 2004,22:16.
  • 6Baltussen E,Sandra P,David F,Cramers C.J.Microcol.Sep.,1999,11:737.
  • 7Frank D,Tienpont B,Sandra P.LCGC North America,2003,21,2:1.
  • 8Wenmin Liu,Hanwen Wang,Yafeng Guan.J.Chromatogr.A 2004,1045:15.
  • 9Arthur C L,Pawliszyn J.Anal.Chem.,1987,62:2145.
  • 10M.Chai,J.Pawliszyn,Environ.Sci.Technol.,1995,29:693.

共引文献20