BacT/Alert240全自动血培养仪的临床应用与评价 被引量:2

Clinical Application and Evaluation of BacT/Alert 240 Automated Blood Culture System
下载PDF
导出
摘要 目的评价全自动血培养系统(BacT/Alert240)的临床应用情况。方法回顾性分析了用BacT/Alert240全自动血培养系统检测的1898份血标本,对检出阳性的时间、细菌种类和阳性率进行了评估分析。结果1898份血标本中分离出255株病原微生物,阳性率为13.4%。最早阳性检出时间为2h,24h内检出的阳性数占72.2%,48h检出的阳性数占92.6%。假阳性率为3.1%,假阴性率为0-3%。结论应用BacT/Alert240全自动血培养系统提高了血培养的阳性率,缩短了阳性检出时间,检出细菌种类多,而且操作简便,结果快速、准确。 Objective To evaluate the clinical application of BacT/Alert 240 automated blood culture system. Methods A total of 1898 blood samples were detected by BacT/Alert 240 automated blood culture system. Positive rate, time to positive, the types of microorganisms recovered were analyzed retrospectively. Results 255 strains were isolated from 1898 blood samples, positive rate was 13.4%. The shortest time to positive was 2h. The cumulative positive rate was 72.2% within 24h, and 92.6% within 48h. False positive rate was 3.1%, and false negative rate was 0.3%. Conclusions The positive detection rate has rised, the detection time to positive reduced.
作者 乔宁 喻华
出处 《世界感染杂志》 2008年第5期361-363,369,共4页 World Journal of Infection
关键词 全自动血培养仪 病原微生物 败血症 检出时间 Automated blood culture system Microorganism Septicemia Time to positive
  • 相关文献

参考文献4

二级参考文献7

共引文献99

同被引文献36

  • 1Kurtz SM,Ong KL,Lau E,et al.Prosthetic joint infection risk after TKA in the Medicare population[J].Clin Orthop Relat Res,2010(468): 52-56.
  • 2Ong KL,Kurtz SM,Lau E,et al.Prosthetic joint infection risk after total hip arthroplasty in the Medicare population[J].Arthroplasty,2009,24(6 Suppl): S105-109.
  • 3Song Z,Borgwardt L,H?iby N,et al.Prosthesis infections after orthopedic joint replacement: the possible role of bacterial biofilms[J].Orthop Rev,2013,5(2): 65-71.
  • 4Lonner JH,Desai P,Dicesare PE,et al.The reliability of analysis of intra-operative frozen sections for identifying active infection during revision hip or knee arthroplasty[J].J Bone Joint Surg Am,1996,78(10): 1553-1558.
  • 5Atkins BL,Athanasou N,Deeks JJ,et al.Prospective evaluation of criteria for microbiological diagnosis of prosthetic-joint infection at revision arthroplasty[J].J Clin Microbiol,1998,36(10): 2932-2939.
  • 6Portillo ME,Salvadó M,Trampuz A,et al.Sonication versus vortexing of implants for diagnosis of prosthetic joint infection.J Clin Microbiol,2013,51(2): 591-594.
  • 7Ince A,Rupp J,Frommelt L,et al.Is aseptic loosening of the prosthetic cup after total hip replacement due to non-culturable bacterial pathogens in patients with low-grade infection[J]? Clin Infect Dis,2004,39(11): 1599-1603.
  • 8Bori G,Soriano A,García S,et al.Low sensitivity of histology to predict the presence of microorganisms in suspected aseptic loosening of a joint prosthesis[J].Mod Pathol,2006,19(6): 874-877.
  • 9Pandey R,Berendt AR,Athanasou NA,et al.Histological and microbiological findings in non-infected and infected revision arthroplasty tissues[J].Arch Orthop Trauma Surg,2000,120(10): 570-574.
  • 10Costerton JW.Bio-film theory can guide the treatment of devicerelated orthopaedic infections[J].Clin Orthop Relat Res,2005(437): 7-11.

引证文献2

二级引证文献18

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部