期刊文献+

闭环式频率适应性起搏器对情绪刺激的反应

Heart rate response during acute mental stress with closed loop stimulation
原文传递
导出
摘要 目的比较情绪反应在起搏器程控为闭环刺激(CLS)和加速度计(DDDR)方式时心房起搏频率的变化。方法共54例患者植入具有CLS和加速度计两种频率适应性算法的起搏器,比较分别将起搏器程控为CLS和DDDR方式时,患者情绪反应对心房起搏频率的影响。结果随访3个月,有35例患者符合情绪反应测试要求,进行情绪反应测试。与将起搏器程控为DDDR方式比较,将起搏器程控为CLS方式时睛绪反应引起的心房起搏频率相关性更好,平均心房起搏频率(70.57±4.80)次/minVS(61.72±3.68)次/min,(P〈0.001)。结论闭环式频率适应性起搏器比体动方式频率适应性起搏器在情绪反应方面具有更好的心房起搏频率调节作用。 Objective The objective of this study was to compare heart rates during emotional response, with the patient's pacemaker programmed to a closed loop stimulation;(CLS) pacing mode and an aceelerometer pacing mode. Methods Fifty-four Patients were implanted with a BIOTRONIK Protos CLS pacemaker (Berlin, Germany) , which equipped with both the CLS technology and an aecelerometer. Rate adaptation to emotional response was achieved through either selecting CLS or the accelerometer sensor. Results CLS provided a statistically significant higher increase in heart rate to emotional response compared with an accelerometer pacing mode (70. 57 ± 4. 80 vs 61.72 ± 3.68, P 〈 0. 001 ). Conclusion the CLS algorithm responds with a more appropriate heart rate response to emotion compared with accelerometer pacing mode.
出处 《中华心律失常学杂志》 2008年第5期361-363,共3页 Chinese Journal of Cardiac Arrhythmias
关键词 情绪反应 闭环刺激方式 体动刺激方式 Emotional response Closed loop stimulation , Accelerometer
  • 相关文献

参考文献5

  • 1Malinowski K. Interindividual comparison of different sensor principles for rate adaptive pacing. PACE, 1998,21:2209-2213.
  • 2Pieragnoli P, Colella A, Michelucci A, et al. A new algorithm for closed-loop stimulation : a feasibility study. PACE, 2003,26:229- 232.
  • 3Coenen M, Malinowski K, Spitzer W, et al. Closed loop stimulation and accelerometer-based rate adaptation: results of the PROVIDE study. Europace,2008,10:327-333.
  • 4Cook L, Tomczak C, Busse E, et al. Impact of a right ventricular impedance sensor on the cardiovascular responses to exercise in pacemaker dependent patients. Indian Pacing Electrophysiol J, 2005,5 : 160-174.
  • 5Chandiramani S, Cohorn LC, Chandiramani S. Heart rate changes during acute mental stress with closed loop stimulation: report on two single-blinded,pacemaker studies. PACE, 2007,30:976-984.

相关作者