摘要
目的探讨化学荧光法(CFEIA)筛查新生儿苯丙酮尿症(PKU)的优越性和制定本地区苯丙氨酸(Phe)筛查切值。方法应用CFEIA法和细菌抑制法(BIA)检测新生儿滤纸干血片上Phe浓度,筛查PKU。结果正常新生儿Phe频数呈正态分布,99%百分位数为116μmol/L(1.93mg/d1),确定本地区筛查切值为1.90mg/dl,荧光法筛查67040人,确诊PKU11例,(1/6096),BIA筛查194012人,确诊PKU 21例(1/9239),经卡方检验,两法无显著性差异(X^2=1.27,P〉0.05);荧光法有6例筛查值在切值附近(2.02~3.72mg/d1),经召回复查分别确诊为轻、中型高苯丙氨酸血症和经典型PKU。结论荧光法定量、灵敏、特异性高,资料永久性保存,是国内PKU筛查值得推荐的方法。
Objective To study the superiority of the chemical fluorometric enzyme immunoassay(CFEIA) in screeing neonatal phenyiketenuria (PKU) and determin the out - off point of phenylalanine (Phe) of neonatal screening for PKU. Methods Phe concentration in dried - blood spot specimens on filter paper was detected by the CFELA or bacterial inhibition assay( BIA). Restults The Phe frequency of normal neonate shows positively distribution. The cut - off point of Phe of neonatal screening for Phe is 1.90 mg/dl in Lianyungang. No difference was found between CFA and BIA (P〉0.05). Among 67 040 neonates screened by CFELA, 11 1/6 096 were found to be defective;194 012 neonates screened by BIA, 21 1/9 239 were found to be defective. Conclusion The guantitative CFELA is more sensitive and rapid,and should be a worthy method being recommend for major screening.
出处
《医学研究杂志》
2008年第10期88-90,共3页
Journal of Medical Research
关键词
化学荧光法
细菌抑制法
苯丙酮尿症
苯丙氨酸
优越性
Chemieal fiuorometrie enzyme immunoassay (CFELA)
Inhibition assoy (BIA)
Phenylnetonurin (PKU)
Phenylalanine (Phe)
Superiority