期刊文献+

数字式多功能外伤清洗机冲洗压力范围的实验研究 被引量:2

Experimental study on the pressure range of digital multifunction trauma rinser
下载PDF
导出
摘要 目的确立数字式多功能外伤清洗机各种喷淋形式的最佳冲洗压力范围。方法选用2~3kg家兔建立浅Ⅱ度烫伤模型、切割伤模型、撕脱伤模型,用大肠杆菌污染创面,10min后分别采用适当喷淋形式冲洗伤口,冲洗后进行伤口大体观察,并取材涂片计数细菌及组织学检查有无二次创伤。结果各种喷淋形式最适合压力范围分别为:莲蓬状水流0.220~0.350MPa,柱状水流0.130~0.250MPa,扁平状水流0.250~0.350MPa,雾状水流0.140~0.300MPa。结论数字式多功能外伤清洗机在适当压力范围内清除细菌及伤口碎屑、碎片或异物的作用更显著,而不损伤正常组织。 Objective To establish the optimal pressure range of digital multifunction trauma rinser (DMTR). Methods Rabbits weighing 2--3 kg were recruited to mimic wounds including superficial second degree scald, concis and avulsed wound, then contamination with Bacillus coli was followed. Irrigating using DMTR was performed in 10 minutes. After evaluating the cleanness of the wounds,bacteria counts in wound were performed through smear method. To judge whether there was concomitant injury or sencond insult,histological examination was done as well. Results Individual appropriate pressure range of sprayers of DMTR was as follows., shower jet 0. 220- 0. 350Mpa; columnar jet 0. 130- 0. 250Mpa; flat jet 0. 250- 0. 350Mpa; pulverization jet 0. 140- 0. 300MPa. Conclusion Irrigating with DMTR in appropriate pressure range can obviously remove germs, fragments and foreign bodies in wound without injurying the normal tissue.
出处 《重庆医学》 CAS CSCD 2008年第22期2564-2565,共2页 Chongqing medicine
关键词 清创术 数字式多功能外伤清洗机 压力范围 治疗 debridement DMTR pressure range therapy
  • 相关文献

参考文献9

  • 1Bahrs C, Schnabel M, Frank T, et al. Lavage of contaminated surfaces., an in vitro evaluation of the effectiveness of different systems[J].J Surg Res, 2003,112 (1):26.
  • 2Leslie T,Aitken DA,Barrie T, et al. Residual debris as a potential cause of postphacoemulsification endophthalmitis[J]. Eye,2003,17(4) : 506.
  • 3Bowler PG,Duerden BI, Armstrong DG. Wound microbiology and associated approaches to wound management [J]. Clin Microbiol Rev, 2001,14(2) : 244.
  • 4王丽姿,李亚洁.创伤伤口冲洗方法及装置的研究进展[J].护士进修杂志,2003,18(8):746-748. 被引量:12
  • 5Hassinger SM, Harding G, Wongworawat MD. High-pressure pulsatile lavage propagates bacteria into soft tissue[J]. Clin Orthop Relat Res, 2005,439 : 27.
  • 6Boyd JI 3rd, Wongworawat MD. High-pressure pulsatile lavage causes soft tissue damage[J].Clin Orthop Relat Res, 2004,427 : 13.
  • 7Adili A,Bhandari M, Schemitsch EH. The biomechanical effect of high-pressure irrigation on diaphyseal fracture healing in vivo[J]. J Orthop Trauma, 2002,16(6) :413.
  • 8Draeger RW, Dahners LE. Traumatic wound debridement:a comparison of irrigation methods[J]. J Orthop Trauma,2006,20(2) :83.
  • 9Polzin B,Ellis T, Dirschl DR. Effects of varying pulsatile lavage pressure on cancellous bone structure and fracture healing[J]. J Orthop Trauma,2006,20(4) :261.

二级参考文献10

共引文献11

同被引文献36

  • 1朱通伯.处理开放性骨折及关节创伤的新观点[J].中华骨科杂志,1995,15(6):393-396. 被引量:77
  • 2郝家骥,方保家.MZ—1冲洗机研制及实验研究[J].中国生物医学工程学报,1996,15(4):365-367. 被引量:2
  • 3Leslie T,Aitken DA, Barrie T, et al. Residual debris as a potential cause of postphacoemulsification endophthalmi- tis[J]. Eye(Lond),2003,17(4) :506-512.
  • 4Bowler PG, Duerden BI, Armstrong DG. Wound microbi- ology and associated approaches to wound management[J]. Clin Microbiol Rev,2001,14(2) :244-269.
  • 5Morse JW, Babson T,Camasso C, et al. Wound infection rate and irrigation pressure of two potential new wound irrigation devices: the port and the eap[J]. Am J Emerg Med,1998,16(1) :37-42.
  • 6Polzin B, Ellis T, Dirschl DR. Effects of varying pulsatile lavage pressure on cancellous bone structure and fracture healing[J]. J Orthop Trauma,2006,20(4) :261-266.
  • 7Adili A, Bhandari M, Schemitsch EH. The biomechanical effect of high-pressure irrigation on diaphyseal fracture healing in vivo[J]. J Orthop Trauma, 2002,16 (6) : 413- 417.
  • 8Draeger RW, Dahners LE. Traumatic wound debridement:a comparison of irrigation methods [J]. J Orthop Trauma, 2006,20(2) :83-88.
  • 9Hassinger SM, Harding G, Wongworawat MD. High-pressure pulsatile lavage propagates bacteria into soft tissue[J]. Clin Orthop Relat Res,2005,439:27-31.
  • 10Boyd JI 3rd. Wongworawat MD. high-pressure pulsatile lavage causes soft tissue damage[J]. Clin Orthop Relat Res,2004(427) :13-17.

引证文献2