期刊文献+

层析γ扫描中的探测效率刻度 被引量:6

Calibration of Detection Efficiency in Tomographic Gamma Scanning
下载PDF
导出
摘要 探测效率刻度技术是层析γ扫描测量中最重要的技术之一。本工作研究用蒙特卡罗方法刻度层析γ扫描系统探测效率的方法。对3×3×3体素组成的样品模型,用蒙特卡罗软件计算了层析γ扫描测量装置的探测效率矩阵。在实验室层析γ扫描原型装置上,实验研究了层析γ扫描测量装置的探测效率。对两者进行了比较,相对偏差绝对值小于5%。研究结果表明了蒙特卡罗方法刻度层析γ扫描测量装置探测效率的可行性。 The calibration technique of detection efficiency is one of the most important techniques in tomographic gamma scanning (TGS) method. The calibration of detection efficiency with Monte-Carlo method was proposed. For a 3×3 ×3 TGS model, all 972 efficiency matrix elements were calculated by Monte-Carlo code. The experimental calibration of detection efficiency in TGS prototype was conducted. The calculated values fit the experimental data very well with relative deviation less than 5%. The results show that Monte-Carlo method used to obtain TGS efficiency is feasible and reliable.
出处 《原子能科学技术》 EI CAS CSCD 北大核心 2008年第11期1043-1047,共5页 Atomic Energy Science and Technology
基金 第二炮兵工程学院科技创新基金资助项目(XY2007JJ06)
关键词 层析γ扫描 探测效率刻度 MONTE-CARLO方法 tomographie gamma seanning detection efficiency calibration Monte-Carlo method
  • 相关文献

参考文献3

  • 1ESTEP R J, PRETTYMAN T H, SHEPPARD G. Tomographic gamma scanning to assay heterogeneous radioactive waste[J]. Nuclear Science and Engineering, 1994, 118: 145-152.
  • 2CAMP D C, MARTZ H E, ROBERSON G P, et al. Nondestructive waste-drum assay for transuranic content by gamma-ray active and passive computed tomography[J]. Nuclear Instruments and Methods in Physics Research A, 2002, 495: 69-83.
  • 3KAK A C, SLANEY M. Principles of computerized tomographic imaging[M]. New York: IEEE Press, 1987.

同被引文献72

  • 1朱文凯,陈军,熊建平,陈坤,陈益军.ISOCS系统无源效率刻度测量方法的准确性检验[J].核电子学与探测技术,2005,25(3):287-290. 被引量:34
  • 2Prettyman T H, Betts S E, Taggart D P, et al. Field Experience with a Mobile Tomographic Nondestructive Assay System. Los Alamos National Laboratory, 1995. LA-UR-95-3501.
  • 3Wang T F, Martz H E, Roberson G P, et al. Three dimensional imaging of a molten-salt-extracted plutonium button using both active and passive gamma-ray computed tomography. Nucl Instrum Methods Phys Res A, 1994, 353:672-677.
  • 4Goodman D, Jackson J A, Martz H E, et al. An Active And Passive Computed Tomography Algorithm With A Constrained Conjugate Gradient Solution. Lawrence Livermore National Laboratory, 1998. UCRL-JC-130818.
  • 5Robert J E, Thomas H P, Sheppard G A. Tomographic gamma scanning to assay heterogeneous radioactive waste. Nucl Sci Eng, 1994, 118: 145-152.
  • 6Venkataraman R, Croft S, Villani M, et al. Performance study of the tomographic gamma scanner for the radioassay of drums. In: Proceedings of 45th Annual INMM Meeting. Orlando: FL, 2004. 18-22.
  • 7Croft S, Venkataraman R, Villani M. Characterizing a tomographic gamma scanner. In: Proceedings of 45th Annual INMM Meeting. Orlando: FL, 2004.18-22.
  • 8Bronson F L, Wang L. Validation of the MCNP Monte Carlo code for germanium detector gamma effieiency calibrations. In: Proceedings of Waste Management'96. Tucson: AZ, 1996.
  • 9Ewa I O B, Bodizs D, Czifrus S, et al. Monte Carlo determination of full energy peak efficiency for a HPGe detector. Appl Radiat Isot, 55(2001): 103-108.
  • 10Selcow E C, Dobrzeniecki A B, Yanch J C, et al. An evaluation of the Monte Carlo simulation of SPECT projection data using MCNP and SimSPECT. In: ICONE 4: ASME/JSME Intemational Conference on Nuclear Engineering, LA: New Orleans, 1996. BNL-62803.

引证文献6