摘要 目的运用16排螺旋CT图像与CARTO三维电解剖标测整合技术(Carto-Merge)指导阵发性房颤的导管消融治疗。方法32例经抗心律失常药物治疗效果欠佳的患者进行手术,其中6例患者术前1~3d行16排CT心脏成像检查,术前将CT导入有Carto-Merge的软件包进行三维重建,重建后在Carto-Merge指导手术。结果所有患者均顺利完成手术,未发生手术并发症。Carto-Merge整合误差为(1.8±0.62)mm,使用Carto-Merge患者平均手术时间为(98±28)min,X线曝光时间为(26±11)min;未使用Carto-Merge患者手术时间为(112±42)min,X线曝光时间为(35±17)min。结论使用图像融合技术可以缩短手术及X线曝光时间,提高手术的效率及安全性。 Objective To investigate the effectiveness and advantages of the Carto-Merge technique in guiding catheter ablation of paroxysmal atrial fibrillation. Methods Thrity-two patients were treated and 6 patients finished 16 slice CT image 1-3d before ablation. 16-slice CT scan combined with Carto image was performed :by using Carto-Merge software and guiding ablation. Results All patients were successfully achieved the endpoint of ablation. All Carto-Merge patients with an error of (1.8 ± 0. 62)mm. The mean procedure time was (98±28)min in Merge group and (112±42)min in no-Merge group. The X-ray exposed time was (26± 11)min in Merge group and (35 ±17)min in no-Merge group. Conclusion Carto Merge technique can shorten the ablation and X- ray exposed time, also improve safety and effectiveness.
出处 《重庆医学》 CAS CSCD 2008年第23期2680-2682,共3页 Chongqing medicine
关键词 导管消融 CT心脏成像 心房颤动 图像融合技术 catheter ablation CT image atrial fibrillation Carto-Merge technique
  • 相关文献

参考文献6

  • 1Marrouche NF, Martin DO, Wazni O, et al. Phase-array intracardiac echocardiography monitoring during pulmo- nary vein isolation in patients with atrial fibrillation: im pact on outcome and eomplieations[J].Cireulation, 2003, 107:2710.
  • 2Sra J,Krum D, Hare J, et al. Feasibility and validation of registration of three-dimensional left atrial models derived from computed tomography with a noncontact cardiac mapping system[J]. Heart Rhythm,2005,2:55.
  • 3Haissaguerre M,Jais P, Shah C, et al. Spontaneous initiation of atrial fibrillation by ectopic beats originating in the pulmonary veins[J]. N Eng J Med, 1998:659.
  • 4Schwartzman D, Lacomis J, Wigginton WG, et al. Characterization of left atrium and distal pulmonary vein morphology using multidimensional computed tomography [J].J Am Coll Cardiol,2003,41 : 1349.
  • 5Kistler PM, Earley M J, Harris S, et al. Validation of three mensiona leardiac image integration: use of integrated CT image into electroanatomic mapping system to perform catheter ablation of atrial fibrillation [J]. J Cardovasc Eletrophysiol, 2006,17 : 341.
  • 6Jun D,Timm D,Darshan D, et al. Initial experience in the use of integrated electroanatomic mapping with three-dimensional MR/CT images to guide catheter ablation of atrial fibrillation[J].J Cardovasc Eletrophysiol, 2006,17 : 459.

同被引文献22

  • 1边昶,马骥,蒋峻,姚述远,吕兴广,王捷.Carto系统指导下典型心房扑动的射频导管消融9例[J].心电学杂志,2009(5):346-347. 被引量:1
  • 2胡大一,孙艺红.心房颤动的流行病学和治疗现状[J].中华全科医师杂志,2006,5(1):5-7. 被引量:85
  • 3李国庆,姚娟,任澎,雷建新,戴晓燕,王钊,阿木提,米日古丽.Carto-Merge技术指导下导管消融阵发性心房颤动的初步探讨[J].临床心血管病杂志,2007,23(6):409-412. 被引量:3
  • 4SUN Y Y,AZAR F S,XU C Y,et al. Registration of high-res- olution 3D atrial images with eleetroanatomical cardiac map- ping:Evaluation of registration methodology [J]. SPIE, 2005, 5744:299-307.
  • 5MIKAELIAN B J, MALCHANO Z J, NEUZIL P, et al. Inte- gration of 3-Dimendinaleardiac computed tomography images with real-time electroanatomie mapping to guide catheter abla- tion of atrial fibrillation[J].Circulation, 2005,112 : e35-e36.
  • 6KISTLER P M,RAJAPPAN K,JAHNGIR M, et al. The im- pact of CT image integration into an electroanatomic mapping system on clinical outcomes ofcatheter ablation of atrial fibril- lation [J]. Journal of Cardiovascular Electrophysiology, 2006, 10(17) :1093-1101.
  • 7BELLA P D,FASSINI G,CIREDDU M, et al. Image Integra- tion Guided Catheter Ablation of Atrial Fibrillation: A pro- spective randomized study [J]. Journal of Clinical Cardiology, 2009,20:258-265.
  • 8ERTAGLIA E,BELLA P D,TONDO C,et al. Image integra- tion increases efficacy of paroxysmal atrial fibrillation catheter ablation: Results from the CartoMerge^TM italian registry [J]. EP-Europace,2009,8(11) : 1004-1011.
  • 9BESL P J, MCKAY N D. A method for registration of 3-D shapes [J]. IEEE Transactions on Pattern Analysis and Ma- chine Intelligence, 1992,14 (2) : 239-256.
  • 10PENNEY G P,EDWARDS P J,KING A P,et al. A Stochastic iterative closest point algorithm (stochastlCP)[J]. MIC- CAI2001,2001,2208: 762-769.