自动夺获和阈值管理功能对起搏器输出电压的影响

Effect of auto-capture feature to ventricular pulse amplitude
下载PDF
导出
摘要 目的探讨自动阈值管理及自动夺获功能对起搏器输出电压的影响,及不同夺获功能原理下起搏器输出电压的差异,以评估自动夺获功能对起搏器寿命的影响。方法分别选取植入自动夺获功能起搏器患者13例和植入无自动夺获功能起搏器患者20例进行临床随访。比较自动夺获功能组起搏器和无自动夺获功能组起搏器输出电压、脉宽等,并比较具有不同阈值夺获原理的起搏器之间输出电压的差异性。结果经统计分析表明具有阈值夺获功能起搏器组平均输出电压为(2.023±1.294)V,无阈值夺获功能起搏器组平均输出电压为(3.622±0.353)V,差异有统计学意义(P<0.01)。两组脉宽并比较差异无统计学意义(P>0.05)。虽然Medtronic公司的Kappa 700系列型起搏器平均起搏电压(2.485V)较ST.JUDE公司的Pacesetter ENTI-TYTM Model DC 5226或5130型起搏器平均输出电压(1.778V)高,但两组比较差异无统计学意义(P>0.05)。结论自动阈值管理或自动夺获功能可有效的减少起搏器输出电压不必要的损耗,节省起搏器电池能量,达到延长起搏器寿命的目的。 Objective To evaluate the influence of auto-capture feature to the ventricular pulse amplitude,and the difference of the ventricular pulse amplitude between two kinds of pacemakers with different auto-capture algorithm. Then,to evaluate the action of auto-capture on the pacemaker longevity theoretically. Methods Thirteen patients implanted pacemakers with auto-capture function and twenty patients without auto-capture function were all followed up. The parameters such as pulse amplitude, pulse width, pacing threshold and the activity auto-capture function were all obtained from the programmer. Then to compare the pulse amplitude and width of these two groups, also do the pacemakers with different auto capture algorithm. Results The mean pacing amplitude of pacemakers with auto-capture was (2. 023±1. 294)V,and that of pacemakers without auto capture was (3. 622±0. 353)V, (P〈 0.01) ,which indicated that there was a significant difference between these two groups. There was no significant difference for pulse width (P〉0.05). Although the mean of pulse amplitude in Kappa 700 series (2. 485V) was higher than the of Pacesetter ENTITYTM model DC 5226 or 5130 (1. 778V) ,there was no difference between that two groups (P〉0.05). Conclusion Capture management or auto capture feature can effectively decrease the consumption of unnecessary pulse amplitude, save the power of battery and prolong the pacemaker longevity.
出处 《重庆医学》 CAS CSCD 2008年第23期2690-2691,2693,共3页 Chongqing medicine
关键词 起搏器 自动阈值管理 输出电压 pacemaker auto-capture management ventricular pulse amplitude
  • 相关文献

参考文献8

  • 1Dodinot B. New pacemaker functions: which ones represent real progress and which are only gadgets[J].Ann Cardiol Angeiol (Paris) ,2000,49(4) :252.
  • 2刘平,王苏加,潘秀荣,王可福,张运.影响埋藏式单腔心脏起搏器寿命相关因素的临床研究[J].山东大学学报(医学版),2004,42(5):558-560. 被引量:3
  • 3Seeger W. Autocapture function was introduced to save battery energy and prolong generator longevity[J]. Pacing Clin Electrophysiol, 1999,22(4 pt 1) :693.
  • 4Suri R, Harthorne JW, Galvin J. Automatically optimizing pacing output:an excellent idea, but with potentially lethal pitfalls[J].Pacing Clin Electrophysiol, 2001,24(4 pt 1) 520.
  • 5杨艳敏,朱俊,宋有城,袁贤奇,谭慧琼,贺丽霞,章友华,李建冬,章晏,刘玉芝.具有自动夺获功能起搏器的临床应用[J].中华心血管病杂志,1998,26(2):120-122. 被引量:2
  • 6Danilovic D, Breivik K, Hoff PI, et al. Clinical performance of steroid-eluting pacing leads with 1. 2mm^2 electrodes[J]. Pacing Clin Electrophysiol, 1997,20(11) : 1799.
  • 7Simeon L,Duru F,Fluri M,et al. The impact of automatic threshold tracking on pulse generator longevity in patient with different chronic stimulation threshold [J].Pacing Clin Electrophysiol,2000,23(11 pt 2) :1788.
  • 8张燕,王景峰,伍卫,聂如琼,周淑娴,孔敏仪.有自动夺获功能的双腔起搏器的临床观察[J].中华心律失常学杂志,2003,7(5):307-308. 被引量:1

二级参考文献4

  • 1Kindermann M, Schwaab B, Berg M, et al. Longevity of dual chamber pacemaker: device and patient related determiants[J]. Pacing Clin Electrophysiol, 2001,24(5):810-815.
  • 2Ribeiro AL, Rincon LG, Oliveira BG, et al. Automatic adjustment of pacing output in the clinical setting[J]. Am Heart J, 2004, 147(1):127-131.
  • 3Schwaab B, Schwerdt H, Heisel A, et al. Chronic ventricular pacing using an output amplitude of 1.0 volt[J].Pacing Clin Electrophysiol, 1997, 20(9):2171-2178.
  • 4Markewitz A, Kronski D, Kammeyer A, et al. Determinants of dual chamber pulse generators longevity [J].Pacing Clin Electrophysiol, 1995, 18 (12):2116-2120.

共引文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部