小动物PET装置的符合系统设计 被引量:3

Coincidence System of Small Animal Positron Emission Tomography Scanner
下载PDF
导出
摘要 符合探测是正电子发射断层扫描仪(PET)的本质,其设计好坏直接影响整个系统的性能。小动物正电子发射断层扫描仪是由中国科学院高能物理所自主研发制造的我国首台动物PET。目前,初步实验结果表明其指标能够应用于核医学的一些基础研究中。本文主要介绍这台动物PET装置的符合系统设计。 Coincidence is one of the most important parts in positron emission tomography, which is crucial to overall performance of the system. A small animal PET has been developed at Institute of High Energy Physics, Chinese Academy of Science, China. In this paper, the coincidence system of the animal PET was introduced. The initial experiment results indicated that the coincidence rate of 40 MHz can be achieved and the system sensitivity can be up to 780 cps/μCi ·μl.
出处 《核电子学与探测技术》 CAS CSCD 北大核心 2008年第5期1012-1015,共4页 Nuclear Electronics & Detection Technology
基金 国家自然科学基金(10475096) 中国科学院仪器研制项目(Y2003012)的支持
关键词 动物PET 符合探测 散点图 Animal PET, coincidence, flood image
  • 相关文献

参考文献9

  • 1Phelps M E, Hoffman E J, Mullani N A, et al.[J]. Nucl Med. 1975, 16:210-24.
  • 2! RowlandDJ, LewisJ S, WelchMJ. J. Cell Biochem. Suppl.2002, 39:110-15.
  • 3Watanabe M, Uchida H, Okada H, et al. IEEE Trans. Med. Iamging 1992, 11:577-80.
  • 4Tai Y, Chatziioannou A, Yang Y, et al. Phys. Med.Biol. 2003, 48:1519-37.
  • 5Knoess C, Siegel S, Smith A, et al.Eur. J. Nucl. Med.Mol.Imaging 2003. 30:737-47.
  • 6Lecomte R, Cadorette J, Rodrigue S, et al. IEEE Trans. Nucl.Sci. 1996. 43:1952-57.
  • 7Dale L B, David W T, Peter E V and Michael N M. Positron Emission Tomography : Basic Science. Springer-Verlag Londom Limited, 2005, p35.
  • 8Casey M E, Nutt R.IEEE Trans. Nucl. Sci. 1986. NS-33(1):460-3.
  • 9Derenzo S E, Budinger T F, Cahoon J L, et al. Nucl. Sci. , 1977, NS24:544-58.

同被引文献24

  • 1何义杰,陈英茂,邵明哲,张书文,尹大一,徐白萱,王瑞民,姚树林,丁勇,关志伟,田嘉禾.能窗下限对三维PET图像质量的影响[J].中华核医学杂志,2004,24(5):305-307. 被引量:3
  • 2周伟,尹端沚,汪勇先.小动物PET[J].核技术,2006,29(3):207-213. 被引量:11
  • 3高飞,刘华锋,施鹏程.小动物正电子断层成像仪性能测试[J].物理学报,2007,56(7):4229-4234. 被引量:2
  • 4Gopal B. Basics of PET Imaging[M]. New York: Springer, 2010.
  • 5单保慈,魏龙.小型PET扫描仪的研制与性能测试[C].首届全国分子核医学暨分子影像学学术交流会资料汇编,2006:54-55.
  • 6Xie QG, Kao CM, Wang X, et al. Potentials of digitally samplingscintillation pulses in timing determination in PET. IEEE TransNucl Sci, 2009 , 56 : 2607-2613.
  • 7Xie QG,Chen YB,Zhu J,et al. Initial implementation of LYSO-PSPMT block detector with an all digital DAQ system. Piscataway :IEEE, 2010: 1759-1762.
  • 8Wang C,Li H,Ramirez RA, et al. A real time coincidence systemfor high count-rate TOF or Non-TOF PET cameras using hybridmethod combining AND-logic and time-mark technology. IEEETrans Nucl Sci, 2010,57; 708-714.
  • 9Tetrault MA, Oliver JF, Bergeron M, et al. Red time coincidencedetection engine for high count rate timestamp based PET. IEEETrans Nucl Sci, 2010,57; 117-124.
  • 10Fontaine R, Belanger F, Viscogliosi N, et al. The hardware andsignal processing architecture of LabPET?,a small animal APD-based digital PET scanner. IEEE Trans Nucl Sci, 2009, 56: 3-9.