医用CT剂量指数测量结果的不确定度评定 被引量:1

Uncertainty analyses of computed tomography dose index
下载PDF
导出
摘要 为了使医用诊断计算机断层摄影装置(CT)X射线辐射源在模体中的剂量指数测量结果的不确定度评定合理,基于用带有100mm长杆电离室的积分剂量仪测量CT剂量指数的数学模型,对其测量不确定度来源进行分析,列出了方差、灵敏系数及标准不确定度的计算公式,从而给出了CT剂量指数测量的合成标准不确定度和扩展不确定度。 In order to make the uncertainty of the dose index measurement in the head phantom reasonable for the medical diagnostic X-ray radiation sourees of the computed tomography (CT), the measurement uncertainty components were analyzed according to the mathematical model for the computed tomography dose index test method with a I00 mm ionization chamber. And the calculated formula for square-error, sensitivity coeflieient and standard deviation were listed in the article. As the results, the standard measurement uncertainty and its expanded uncertainty of CT dose index were presented.
作者 高铁 李国琛
出处 《中国测试技术》 2008年第6期21-23,43,共4页 CHINA MEASUREMENT & TESTING TECHNOLOGY
关键词 医用诊断计算机断层摄影装置(CT)X射线辐射源 剂量指数 测量不确定度 Medical diagnostic X-ray radiation source for computed tomography Dose index Uneertainty