期刊文献+

GD-300系列辐射光致发光玻璃剂量计的辐射剂量学特性 被引量:8

Dosimetric Properties of GD-300 Series of Radiophotoluminescent Glass Detector
下载PDF
导出
摘要 为掌握新引进的辐射光致发光玻璃剂量计的辐射剂量学特性,通过用次级标准剂量学实验室的X射线照射装置、60Coγ放射源(1250 keV)及经检定的137Csγ放射源(662 keV)照射,考察了GD-300系列辐射光致发光玻璃剂量计的剂量线性、均匀性、重复性、衰退特性和能量响应等剂量学性能。实验研究结果表明,GD-300系列辐射光致发光玻璃剂量计的辐射剂量学性能良好,适合作为放射工作人员外照射剂量、放射诊疗受检者与患者剂量的测量,也适用于环境辐射的长期累积测量。 In order to grasp the dosimetric toluminescent glass dosimeter (RPLGD), properties of the newly imported radiophoexperiments for studying the dose linearity, uniformity and reproducibility of signal, fading effect and energy dependence of the RPLGD were performed by using the X-ray emitting device and 60Co as well as 137Cs sources in a secondly standard dosimetry laboratory. The general dosimetric properties of the RPLGD are fairly good. It indicates that the RPLGD is suitable for monitoring the external exposure of radiation workers, evaluating the patient dose from medical exposure, and monitoring the long-term integrated environmental radiation levels.
出处 《原子能科学技术》 EI CAS CSCD 北大核心 2008年第12期1120-1124,共5页 Atomic Energy Science and Technology
基金 国家自然科学基金资助项目(10675036)
关键词 辐射光致发光 剂量线性 均匀性 重复性 衰退 radiophotoluminescence dose linearity uniformity reproducibility fading effect
  • 相关文献

参考文献12

  • 1MOSCOVITCH M, JOHN T J S, CASSATA J R, et al. The application of LiF: Mg, Cu, P to large scale personnel dosimetry: Current status and future directions[J]. Radiat Prot Dosim, 2006, 119(1-4) : 248-254.
  • 2郑钧正.国产氟化锂片型热释光探测器性能比较研究[J].辐射防护,1997,17(3):225-231. 被引量:10
  • 3BEKER K. High T-dose response of recent silveractivated phosphate glasses [J ]. Health Phys, 1965, 11(6): 523-529.
  • 4PIESCH E, BURGKHARDT B, VILGIS M. Photoluminescence dosimetry-progress and state of art[J]. Radiat Prot Dosim, 1990, 33 (1-4):215-226.
  • 5OBERHOFER M,SCHARMANN A.应用热释光剂量学[M].北京:中国计量出版社,1988:118-120.
  • 6AKSELROD M S, KORTOV V S, KRAVETSKY D J, et al. Highly sensitive thermoluminescent anion-defective α-Al2O3 :C single crystal detector[J]. Radiat Prot Dosim, 1990, 32.. 15-21.
  • 7YUKIHARA E G, MITTANI J C, VANHAVERE F, et al. Development of new optically stimulated luminescence neutron dosimeters [C/OL]//15th International Conference on Solid State Dosimetry, Delft, The Netherlands, July 8-13, 2007: international solid state dosimetry organization. [2007-08-05]. http://www.ssd15.org/abstracts/2-06.doc.
  • 8YASUDA H, FUJITAKA K. Solid-state integrating detectors as an indicator of biological doses from HZE particles[J]. Advances in Space Research, 2002, 30(4): 927-932.
  • 9ARAKI F, MORIBE N, SHIMONOBU T, et al. Dosimetrie properties of radiophotoluminescent glass rod detector in high-energy photo beams from a liner accelerator and Cyber-Knife [J]. Med Phys, 2004, 31(7): 1 980-1 985.
  • 10吴毅 王时进 郑钧正 等.估算X线诊断受检者器官剂量的体模实验方法.辐射防护,1987,7(5):335-340.

二级参考文献7

共引文献12

同被引文献94

引证文献8

二级引证文献40

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部