摘要
本文研究了比浊法测定含铜基体样品中微量Cl^-的快速分析方法。方法液体最低检出浓度为0.2mg·L^-,固体最低检出浓度为5μg·g^-1。测定含Cl^-为0.0014%,0.0058%,0.0070%的CuSO4,CuCO3·Cu(OH)2和CuO样品,相对标准偏差分别为4.5%,3.2%,2.4%;加标回收率分别为98.2%-105.1%,98.8%~107.6%,96.4%-105.5%。
This paper studied a fast analysis method of detecting micro- amount chloride ions in cupric matrix with turbidimetry, with 0.2mg·L^-1 liguid detection limit and 5μg·g^-1 solid detection limit. As for different samples of cupric sulfate ( chloride content 0. 0014% ) and cupric carbonate basic ( chloride content 0. 0058% ) and cupric oxide ( chloride content 0.0070% ), the respective relative standard deviation are 4.5%, 3. 2% and 2.4%. The recovery rate for the sample added with standard are 98.2%- 105.1%, 98.8%-107. 6%, and 96.4%-105.5%.
出处
《化学工程师》
CAS
2008年第9期37-39,共3页
Chemical Engineer
关键词
比浊法
微量氯
含铜基体样品
测定
turbidimetry
micro- amount chloride ion
cupric matrix