摘要
实验研究了电荷注入检测器(CID)-端视电感耦合等离子体(AVICP)光谱中分析动态范围,测量了Al、B、Be、Cd、Cu、Co、Fe、Li、Sr、V、K、Sr等12个元素的不同光谱级次的44条分析工作曲线。结果表明,分析动态范围为4~5个数量级,无明显自吸收,Lomakin公式的自吸收系数b=1,说明在CID-AVICP光谱中有较宽的分析动态范围。
In this paper the analytical linear ranges of calibration graphs for a new type emission spectrometer with two dimensional charge injection detector and axially viewed plasma were investigated. Calibration graphs for Al, Cd, Cu, Co, Fe, Li, Sr, V, K, Sr, Be, B in different spectral order were examined. The results showed that linear ranges are over four orders of magnitude and no self-reversal or self-absorption is found under concentration at 100-1000 mg/L. The self-absorption coefficient b for Lomakin equation has been determinad and the value is approximately equal to 1.
出处
《分析化学》
SCIE
EI
CAS
CSCD
北大核心
1998年第1期44-47,共4页
Chinese Journal of Analytical Chemistry
关键词
电感耦合
等离子体光谱
分析动态范围
CID
Inductively coupled plasma-atomic emission spectrometry, axially viewed plasma, charge injection array detector, analytical dynamic range