鼻咽癌常规放疗中面颈联合野和下颈野衔接的探讨 被引量:5

The junction of the facio-upper neck field and the tangential low neck-supraclavicular field in conventional radiotherapy for nasopharyngeal carcinoma
下载PDF
导出
摘要 目的:比较鼻咽癌常规放射治疗中面颈联合野和下颈切线野不同衔接方法及间距的衔接平面的剂量分布。方法:利用CMS计划系统来计算面颈联合野和下颈切线野在机架角0度方向上(即颈前部)共线衔接、间隔5mm衔接、重叠1mm和2mm衔接以及机架角在90度或270度方向上(即颈侧方向上)共线衔接平面的剂量分布。结果:两者在0度方向上共线衔接、间隔5mm、重叠1mm和2mm衔接以及在90度或270度方向上共线衔接时,衔接平面最高剂量分别是参考剂量的103%、54%、114%、124%、73%,而在0度方向上共线衔接、重叠1mm和2mm衔接平面的95%剂量曲线分别位于皮下2.5cm、0.8cm、0.4cm。结论:鼻咽癌常规放射治疗中面颈联合野和下颈切线野全束照射时推荐在0度方向上1mm的重叠衔接。 Objective:To compare the dose distribution at the match-plane of varies the junction methods and fields gap of the facio-upper neck field and the tangential low neck-supraclavicular field in conventional radiotherapy for nasopharyngeal carcinoma (NPC). Methods:It was observated that the dose distribution at the match-plane of mono-line junction in the 0 degree direction, 5mm intervals, 1 mm and 2mm overlap, mono-line junction in the 90 degree direction of the facio-upper neck field and the tangential low necksupraelavicular field by three dimensions treatment planning system. Results:Maximun dose at the match-plane of mono-line junction in the 0 degree direction, 5mm intervals, 1 mm and 2mm overlap, mono-line junction in the 90 or 270 degree direction of the two fields were 103% ,54%, 114%, 124% ,73% of the weight dose,respectively; 95% isodose line at the match-plane of mono-line junction in the 0 degree direction, 1 nun and 2mm overlap of the two fields lied 2. 5cm, 0. 8cm,0. 4cm under the surface of skin, respectively. Conclusion: One millimeter overlap junction in the 0 degree direction of the two fields is recommended in the practical work.
出处 《临床肿瘤学杂志》 CAS 2008年第12期1108-1110,共3页 Chinese Clinical Oncology
关键词 鼻咽肿瘤 放射疗法 照射野衔接 Nasopharyngeal carcinoma Radiotherapy Irradiation field junction
  • 相关文献

参考文献13

  • 1肖光莉,蔡伟明,徐国镇,胡郁华,吴雪林,秦德兴,严洁华,李素艳,高黎,肖建平,王纹.CT上的咽旁间隙异常对鼻咽癌T分期的影响[J].中华放射肿瘤学杂志,1993,2(1):4-5. 被引量:67
  • 2Kam MK, Chau RM, Suen J, et al. Intensity-modulated radiotherapy in nasopharyngeal carcinoma: dosimetric advantage over conventional plans and feasibility of dose escalation [ J ]. Int J Radiat Oncol Biol Phys, 2003,56 : 145 - 157.
  • 3吴超权,钟鹤立,李先明,李而周,胡岳然,陈善义.鼻咽癌放射治疗照射野最佳衔接方法的探讨[J].中华放射肿瘤学杂志,2002,11(2):88-90. 被引量:21
  • 4Kron T, Barnes K, O'Brien P. Multicentre dosimetric comparison of photon-junetioning techniques in head and neck radiotherapy [ J ]. Australas Radiol, 2003,47:289 - 294.
  • 5Zhu L, Kron T, Barnes K, et al. Junctioning of lateral and anterior fields in head and neck cancer: a dosimetric assessment of the monoisocentric technique ( including reproducibility) [J]. Int J Radiat Oncol Biol Phys, 1998,41:227 -232.
  • 6Sohn JW, Suh JH, Pohar S. A method for delivering accurate and uniform radiation dosages to the head and neck with asymmetric collimators and a single isocenter[ J ]. Int J Radiat Oncol Biol Phys, 1995,32:809 -813.
  • 7赵路军,傅卫华,戴建荣,高黎,张红志,徐国镇.鼻咽癌半束照射的剂量分布[J].中华放射肿瘤学杂志,2003,12(3):183-187. 被引量:10
  • 8Saw CB, Krishna KV, Enke CA, et al. Dosimetric evaluation of abutted fields using asymmetric collimators for treatment of head and neck [ J ]. Int J Radiat Oncol Biol Phys, 2000,47:821 - 824.
  • 9Saw CB, Hussey DH. Dosimetric assessment of nonperfectly abutted fields using asymmetric collimators [ J ]. Med Dosim, 2000,25:23 - 26.
  • 10钟鹤立,夏俊贤,刘剑锋,李隆兴,李晓东.独立准直器位置精度对半野联合处剂量的影响[J].中国医学物理学杂志,2004,21(2):70-72. 被引量:4

二级参考文献19

  • 1严洁华 谷铣之 等.鼻咽癌的放射治疗.肿瘤放射治疗学[M].北京:北京医科大学中国协和医科大学联合出版社,1993.452-462.
  • 2闵华庆 张天泽 等.鼻咽癌的治疗.肿瘤学[M].天津:天津科学技术出版社,1996.1083-1086.
  • 3钱立庭 陈振东 等.放射性周围神经损伤.实用肿瘤并发症诊断治疗学[M].合肥:安徽科学技术出版社,1997.260-262.
  • 4戴建荣 胡逸民.不对称射野和多叶准直器射野处方剂量计算.肿瘤放射物理学[M].北京:原子能出版社,1999.197.
  • 5孙仲藩 孙庆选 见:徐燮渊 俞受程 曾荻闻 等主编.鼻咽癌[A].见:徐燮渊,俞受程,曾荻闻,等主编.现代肿瘤放射治疗学[C].北京:人民军医出版社,2000.404.
  • 6Zhu L, Kron T, Barnes K, et al. Juctioning of lateral and anterior fields in head and neck cancer: a dosimetric assessment of the monoisocentric technique ( including reproducibility ). Int J Radiat Oncol Biol Phys,1998,41:227-232.
  • 7Sohn JW, Suh JH, Pohar S. A method for delivering accurate and uniform radiation dosages to the head and neck with asymmetric collimators and a single isocenter. Int J Radiat Oncol Biol Phys, 1995,32:809-813.
  • 8Saw CB, Krishna KV, Enke CA, et al. Dosimetric evaluation of abutted fields using asymmetric collimators for treatment of head and neck. Int J Radiat Oncol Biol Phys, 2000,47:821-824.
  • 9Rosenthal DI, Medonough J, Kassaee A. The effect of independent collimator misalignment on the dosimetry of abutted half-beam blocked fields for the treatment of head and neck cancer. Radiother Oncol,1998,49 : 273-278.
  • 10Fabrizio PL, Mccullough EC, Foote RL. Decreasing the dosimetric effects of misalignment when using a mono-isocentric technique for irradiation of head and neck cancer. Int J Radiat Oncol Biol Phys,2000,48:1623-1634.

共引文献121

同被引文献37

引证文献5

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部