熵指数监测全麻患者镇静水平的准确性 被引量:4

Accuracy of spectral entropy as an electroencephalographic measure of level of sedation during general anesthesia
原文传递
导出
摘要 目的评价反应熵和状态熵监测全麻患者镇静水平的准确性。方法择期行腹部手术患者20例,ASAⅠ或Ⅱ级,人室后监测反应熵(RE)、状态熵(SE)及脑电双频谱指数(BIS),静脉注射异丙酚、维库溴铵和芬太尼麻醉诱导,气管插管后机械通气,吸人七氟烷、间断静脉注射维库溴铵和芬太尼维持麻醉。分别于入室时、意识消失前10min、意识消失即刻、气管插管时、手术1h、意识恢复前10min、意识恢复即刻、拔管后10min时记录RE、SE和BIS。结果RE、SE和BIS在意识改变前后差异均有统计学意义(P〈0.05),RE、SE和BIS判断意识消失的临界值分别为76、73和68,灵敏度分别为94%、95%和92%,特异度分别为92%、94%和91%,临界值判断意识消失的准确度分别为93%、95%、94%;判断意识恢复的临界值分别为82、75和70,灵敏度分别为95%、95%和91%,特异度分别为93%、96%和93%,临界值判断意识恢复的准确度分别为98%、96%和97%。结论熵指数能够准确地监测全麻患者镇静水平。 Objective To evaluate the accuracy of spectral entropy including response entropy (RE) and state entropy (SE) as a new electroencephalographic measure of depth of sedation during general anesthesia. Methods Twenty ASA Ⅰ or Ⅱ patients aged 20-55 yr undergoing elective abdominal surgery were studied. Anesthesia was induced with intravenous propofol, fentanyl and vecuronium. After tracheal intubation anesthesia was maintained with sevoflurane inhalation and intermittent iv boluses of fentanyl and veeuronium. BP, HR, SpO2, spectral entropy (Datex-Ohmeda) and BIS (Aspect Medical) were monitored during anesthesia. Loss of eyelash reflex and unresponsiveness to verbal stimulus and mild prodding and shaking were considered to be signs of loss of consciousness (LOC). The time of regaining consciousness was defined as the time when patients started responding to verbal commands. RE, SE and BIS were recorded before anesthesia ( baseline), 10 rain before LOC, immediately after LOC, immediately after tracheal intubation, 1 h after skin incision, 10 min before regaining consciousness immediately after consciousness was regained and 10 min after extubation. Results There were significant changes in RE, SE and BIS during the transition from consciousness to LOC. The RE, SE and BIS values for LOC were 76, 73 and 68 with sensitivity of 94%, 95% and 92% and specificity of 92%, 94% and 91% respectively. The accuracy of RE, SE and BIS in predicting LOC was 93%, 95% and 94% respectively. The RE, SE and BIS values for regaining consciousness were 82, 75 and 70 with sensitivity of 95 %, 95 % and 91% and specificity of 93%, 96% and 93% respectively. The accuracy in predicting regaining of consciousness was 98 %, 96 % and 97 % respectively. Conclusion Entropy index can accurately monitor level of sedation during general anesthesia.
出处 《中华麻醉学杂志》 CAS CSCD 北大核心 2008年第11期1013-1015,共3页 Chinese Journal of Anesthesiology
基金 国家自然科学基金资助项目(30340075)
关键词 脑电描记术 清醒镇静 麻醉 全身 监测 生理学 Electroencephalography Conscious sedation Anesthesia, general Monitoring, physiologic
  • 相关文献

参考文献4

  • 1Ellerkmann RK, Liermaan VM, Alves TM, et al. Spectral entropy and bispectral index as measures of the electroencephalographic effects of sevoflurane. Anesthesiology, 2004, 101 : 1275-1282.
  • 2Vanluchene AL, Vereecke H, Thas O, et al. Spectral entropy as an electroencephalograpbie measure of anesthetic drug effect: a comparison with bispectral index and processed midlatency auditory evoked response. Anesthesiology, 2004,101:34-42.
  • 3Vanluchene AL, Struys MM, Heyse BE, et al. Spectral entropy measurement of patient responsiveness during propofal and remifentanil. A comparison with the bispectral index. Br J Anaesth, 2004, 93: 645-654.
  • 4毕素萍,张宏.熵指数用于全身麻醉脑电监测[J].军医进修学院学报,2007,28(4):308-310. 被引量:3

二级参考文献20

  • 1ViertiO-Oja H,Maja V,Sarkela M.Description of the Entropy algorithm as applied in the Datex-Ohmeda S/5 Entropy Module[J].Acta Anaesthesiol Scand,2004,48:154-161.
  • 2Paloheimo M.Quantitative surface electromyography(qEMG):applications in anaesthesiology and critical care[J].Acta Anaesthesiol Scand Suppl,1990,93:1-83.
  • 3Struys M,Versichelen L,Mortier E.Comparison of spontaneous frontal EMG,EEG power spectrum and bispectral index to monitor propofol drug effect and emergence[J].Acta Anaesthesiol Scand,1998,42:628-636.
  • 4Tammisto T,Toikka O.Spontaneous EMG activity for detection of arousal during general anaesthesia:Comparison between recordings from frontal and neck musculature[J].Eur J Anaesthesiol,1991,8:109-114.
  • 5Bruhn J,Bouillon TW,Shafer SL.Onset of propofol-induced burst suppression may be correctly detected as deepening of anaesthesia by approximate entropy but not by bispectral index[J].Br J Anaesth,2001,87:505-507.
  • 6Sarkela M,Mustola S,Seppanen T.Automatic analysis and monitoring of burst suppression in anesthesia[J].J Clin Monit Comput,2002,17:125-134.
  • 7Johansen JW.Esmolol promotes electroencephalographic burst suppression during propofol/alfentanil anesthesia[J].Anesth Analg,2001,93:1526-1531.
  • 8Bruhn J.Burst suppression ratio is the only determinant for BIS values below 30[J].Can J Anaesth,2002,49:755-756.
  • 9Vakkuri A,Hankala-Yli,Talja P.Time-frequency balanced spectral entropy as a measure of anesthetic drug effect in central nervous system during sevoflurane,propofol,and thiopental anesthesia[J].Acta Anaesthesiol Scand,2004,48:145-153.
  • 10Bruhn J,Bouillon TW,Shafer SL.Bispectral index (BIS) and burst suppression:Revealing a part of the BIS algorithm[J].J Clin Monit Comput,2000,16:593-596.

共引文献2

同被引文献29

引证文献4

二级引证文献8

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部