摘要
The wall electrode sensors have been widely used in the measurement of transport process.However,the theoretical equations in prior work did not include the effect of mass transfer to the edge of wall probe.This effect will bring an important contribution to the electrode current if the probe is small enough,which a lager measuring error if the original equation is used.This paper gives some modified equations in which the effect of mass transfer to the probe edge is considered.The experiment results pressented have supported these equations.
The wall electrode sensors have been widely used in the measurement of transport process.However,the theoretical equations in prior work did not include the effect of mass transfer to the edge of wall probe.This effect will bring an important contribution to the electrode current if the probe is small enough,which a lager measuring error if the original equation is used.This paper gives some modified equations in which the effect of mass transfer to the probe edge is considered.The experiment results pressented have supported these equations.
出处
《化工学报》
EI
CAS
CSCD
北大核心
1998年第2期225-228,共4页
CIESC Journal
基金
国家自然科学基金!No.29576262
关键词
边缘传质
流场参数测试
壁面电极传感器
mass transfer to the edge of electrode,electrochemical measurement,measurement of flow parameter