元宫型宫内节育环对盆腔放射治疗剂量分布的影响 被引量:3

Impact of intrauterine contraceptive ring on pelvic radiotherapeutic dosage distribution
下载PDF
导出
摘要 目的测量元宫型金属宫腔内节育环对放射治疗引起的密度效应及次级射线引起的散射效应对放射治疗剂量分布的影响。方法将元宫型金属节育环置于厚20cm的人体等效组织密度有机玻璃中间,分别用6MV和10MVX线垂直或平行于节育环分别进行10cm×10cm单野照射和二野对穿照射,用FARMER剂量仪电离室测量并比较节育环前1cm和0.5cm,环中间0cm,环后0.5cm和1cm及旁开0.5cm处的剂量。结果垂直或平行于节育环进行10cm×10cm单野照射时,各点与无环时比较,最大剂量变化为0.15%;对穿野照射时,最大剂量变化为0.13%。结论宫内节育器对放射治疗时剂量分布的影响远小于5%,在盆腔肿瘤需要放射治疗时,可以忽略节育环对剂量分布的影响。 Objective To determine the scattering and density effects of intrauterine contraceptive ring on pelvic radiotherapeutic dosage distribution. Methods A contraceptive ring was set into the middle of 20 cm thick plexiglass, an iso-human tissue, and irradiated in 10 cm × 10 cm single field and two parallel fields by 6 MV X ray as well as 10 MV X ray. The ionization measurement was performed by FARMER chamber, and measurement points were chosen in front of ring (0.5 and 1 cm) , in the centre of ring, behind ring (0.5 and 1 cm) , and on the side of ring (0.5 cm) in the presence or absence of ring. Results When 10 cm × 10 cm single field irradiated, the maximal dosage deviation was 0. 15% compared to that without ring. When two parallel fields irradiated, the dose maximal dosage deviation was 0. 13%. Conclusion The dosage change caused by scattering and density effects of contraceptive ring is far less than clinically limited 5% dosage deviation. The effects of contraceptive ring on pelvic radiotherapy can be ignored.
出处 《上海交通大学学报(医学版)》 CAS CSCD 北大核心 2009年第1期61-63,共3页 Journal of Shanghai Jiao tong University:Medical Science
关键词 盆腔 节育环 放射治疗剂量 pelvic cavity intrauterine contraceptive ring radiotherapeutic dosage
  • 相关文献

参考文献8

二级参考文献20

  • 1李先明,胡逸民,哈思衡,张红志,张春利,张信,徐国镇.气腔对高能X线剂量分布的影响[J].中华放射肿瘤学杂志,1992(3):52-56. 被引量:8
  • 2翟仁友,戴定可.食管支架治疗食管良恶性狭窄(附23例分析)[J].中华放射学杂志,1995,29(7):465-468. 被引量:146
  • 3王争,中华放射肿瘤学杂志,1993年,2期,188页
  • 4谷铣之,肿瘤放射治疗学,1983年,14页
  • 5Indra J.Das, Faiz M. Kahn. Backscatter dose perturbation at high atomic number interfaces in megavoltage photon beams[J]. Med. Phys.,1989,16(3):37-375.
  • 6Tony Y Eng. Dose attenuation through a titanium alloy hip prosthesis[J]. Medical dosimerty, 1999,25(1):7-8.
  • 7X. ALLEN LI et al. Radiotherapy dose perturbation of metallic esophageal stents[J]. Int J Radiation Oncology Biology Physics, 2002,54(4): 1276-1285.
  • 8Lin SY et al. The effect of a metal hip prosthesis on the radiatio n dose in therapeutic photon beam irradiations[J]. Appl Radiat Isot, 2002,57(1): 17-23.
  • 9Rebort et al. How accurate is a CT-based dose calculation on a pencil beam TPS for a patient with a metallic prosthesis [J]. Phys.Med Biol., 2001,46(2001):N227-N234.
  • 10Chester Reft et al. Dosimetric consideration for patients with HIP prostheses undergoing pelvic irradiation. Report of the AAPM Ra diation Therapy Committee Task group 63[R]. Med. Phys., 2003,30 (6):1162-1182

共引文献80

同被引文献19

引证文献3

二级引证文献11

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部