期刊文献+

停流─—反相流动注射分析法测定环境水样的化学需氧量 被引量:11

DETERMINATION OF CHEMICAL OXYGEN DEMAND IN AQUEOUS ENVIRENMENTAL SUMPLES BY STOP-REVERSE FLOW^INJECTION METHOD
下载PDF
导出
摘要 以铈(IV)作氧化剂,停流——反相流动注射分析法快速测定环境水样的化学需氧量。方法测定范围2.5~160mg/L,采样频率12h-1。检测下限和标准偏差分别为2.5mg/L和0.9%,cl-至2000mg/L无明显干扰,与经典的重铬酸钾法相对照,结果相关性良好。试剂费用节约95%以上,分析周期缩短至1/24且污染小。 A method using cerium (IV) sulphate as oxidizing agent is described for the determination of chemincal oxygen demand (COD) in aqueous environmental samples by stop-reverse flow-injection with the range of 7. 5- 160mg/L. COD values for sample correlate well with those obtained by standard manual method, except for higher frequency.Less polution and more economical costs. The detection limit and relative standard deviation of the method are 2. 5 mg/L and 0. 9% respectively. Chroride ion up to the levels of 2000 mg/L does not interfere 1 and a sumple throughout of 12h-1 can be achieved.
作者 全晓塞 邵珍
机构地区 宁夏大学
出处 《宁夏农学院学报》 1998年第1期34-36,共3页 Journal of Ningxia Agricultural College
关键词 化学需氧量 流动注射分析 光度法 环境水样 bhemical oxygen demand folw injection analysis cerium (IV ) spectrophotometry
  • 相关文献

参考文献1

二级参考文献12

  • 1徐燕军,Anal Chim Acta,1994年,292卷,191页
  • 2徐燕军,Anal Lett,1994年,27卷,793页
  • 3Gao T,Talanta,1993年,40卷,1927页
  • 4Le X C,Appl Organomet Chem,1992年,6卷,161页
  • 5金钦汉,岩矿测试,1992年,11卷,1/2期,87页
  • 6郭振库,分析化学中的微波制样技术.原理及应用,1992年
  • 7王--,痕量分析,1989年,5卷,1期,56页
  • 8郎春燕,理化检验,1989年,25卷,5期,59页
  • 9金钦汉,分析化学,1988年,16卷,7期,668页
  • 10吴瑞林,丁曼蓉.微波能加热及其在化学分析中的应用[J].分析试验室,1991,10(2):51-54. 被引量:10