带瓣膜主动脉瓣支架的研制及经导管植入动物实验 被引量:6

Preparation of valved aortic stent and trans-catheter implantation to ascending aorta in vivo
下载PDF
导出
摘要 目的:研制可经导管植入的带瓣膜主动脉瓣支架,探索经导管植入该瓣膜支架的可行性。方法:采用0.2mm镍钛记忆合金丝编织内径为15~19mm的网状支架。取牛心包,经固定处理后,按仿形类比设计的方法修剪成瓣叶,并缝合在支架的远端,制成带瓣膜主动脉瓣支架。所有瓣膜支架行体外冲水实验、释放实验、体外管道内释放及静态实验。小香猪7只,采用经导管逆行的方法行升主动脉内置入瓣膜支架的动物实验。结果:研制的带瓣膜主动脉瓣支架,经体外试验证明瓣膜关闭良好,反向均为中心水流,并明确该瓣膜支架能经导管顺利释放,释放后能迅速完全扩张,在管道内固定牢靠。采用12F~16F的输送导管,6只实验动物经髂总动脉植入,所有瓣膜支架均能顺利跨过主动脉弓达升主动脉,并顺利释放,除1例瓣膜支架释放后未完全扩张、移位外,其余均迅速完全扩张、固定良好。24h后处死动物解剖,见支架固定,支架内未见血栓。结论:该带瓣膜主动脉瓣支架启闭功能、自膨功能良好,压缩后的体积符合经导管植入的要求,可采用逆行的方法经髂总动脉植入于升主动脉内,释放后能迅速扩张,固定牢靠。 Objective:To develop valved aortic stent and to assess the feasibility of its trans-catheter implantation to the ascending aorta in vivo. Methods: We designed self-expandable nitinol stents (15-19 mm in diameter) according to the internal diameter of the ascending aorta of Chinese miniswines. Bovine pericardium was cross-linked with 0. 6 % glutaraldehyde solution for 36 hours, and then was shaped and sutured into the stents. Fluid passing test, pre-releasing test and static test of pressure in the tube were performed in all devices. The devices were tried to be inserted percutaneously through common iliac artery to deploy in the ascending aorta in 7 Chinese miniswines. Results: In vitro test showed that the closure of the percutaneous valved aortic stent leaflets was satisfactory, and the fluid flow was not restricted in the opposite direction. The devices could be released through the catheter, expanded completely, and be fixed rapidly in the tube. The devices were deployed using 12F-16F catheter and were inserted through common iliac artery in 6 of the 7 Chinese miniswines. All devices were implanted in the desired position, with 6 expanded completely and fixed rapidly in the ascending aorta. One did not expand completely and was displaced. Animals were executed and dissected after 24 hours. All the devices were fixed in the ascending aorta and there were no tbrombus. Conclusion: The function of the percutaneous valved aortic stent is satisfactory. The size of compressed device is suitable for catheterization. It can be successfully implanted into the ascending aorta by using a retrograde method through common iliac artery.
出处 《第二军医大学学报》 CAS CSCD 北大核心 2009年第2期120-123,共4页 Academic Journal of Second Military Medical University
基金 教育部科学技术研究重点项目(0210)~~
关键词 主动脉瓣 支架 心脏瓣膜假体植入 心脏导管插入术 valved aortic stents heart valve prosthesis implantation heart catheterization
  • 相关文献

参考文献13

  • 1Andersen H R,Knudsen L L, Hasenkam J M. Transluminal implantation of artificial heart valves: description of a new ex pandahle aortic valve and initial results with implantation by catheter technique in closed chest pigs[J]. Eur Heart J. 1992, 13,704- 708.
  • 2Cribier A. Eltchaninoff H. Bash A. Borenstein N, Tron C, Bauer F. et al. Percutaneous transcatheter implantation of an aortic vah'e prosthesis for calcific aortic stenosis: first human casedescription[J]. Circulation, 2002, 106:3006- 3008.
  • 3WebbJ G.Lichtenstein ,S. Transcatheter percutaneous and transapical aorlic valve replacement[J]. Semin Thorac Cardiovasc Surg. 2007. 19:304-310.
  • 4Eltchaninoff H, gajarias A. Tron C.Litzler P Y, Baala B, Godin M, et al. Transcatheter aortic valve implantation:technical aspects, results and indications[J]. Arch Cardiovasc Dis, 2008, 101:126- 132.
  • 5Grube E, Schuler G, Buellesfeld L. Gerckens U, Linke A, Wenaweser P, et al. Percutaneous aortic valve replacement for severe aortic stenosis in high-risk patients using the second- and current third generation selfexpanding CoreValve prosthesis: device success and 30-day clinical outcome[J]. J Am Coll Car diol, 2007, 50:69- 76.
  • 6Feldman T. Percutaneous valve repair and replacement: challenges encountered, challenges met, challenges ahead[J]. Circulation, 2006, 113:771-773.
  • 7宗刚军,白元,秦永文,姜海滨,赵仙先,邹毅清.带瓣膜二尖瓣支架的研制及体外经导管植入实验[J].第二军医大学学报,2008,29(2):197-199. 被引量:3
  • 8白元,宗刚军,秦永文,王洪如,姜海滨.带瓣膜三尖瓣支架的研制及体外经导管植入实验[J].第二军医大学学报,2008,29(2):200-202. 被引量:2
  • 9Vahanian A, Alfieri O R, AI Attar N, Antunes M J, Bax J, Cormier B, et al. Transeatheter valve implantation for patients with aortic stenosis: a position statement from the European Association of Cardio-Thoracic Surgery (EACTS) and the European Society of Cardiology (ESC), in collaboration with the European Association of Percutaneous Cardiovascular Interventions (EAPCI) [J]. Eur J CardiothoracSurg, 2008, 34:1- 8.
  • 10Webb J G, Chandavimol M, Thompson C R, Ricci D R, Carere R G, Munt B I, et al. Percutaneous aortic valve implantation retrograde from the femoral artery[J]. Circulation, 2006, 113 : 842- 850.

二级参考文献10

  • 1申彪,付明森,郭志坤,马建军,陈志国,张圣旭,陈现杰.山羊心脏二尖瓣复合体的应用解剖学研究[J].解剖科学进展,2005,11(2):164-166. 被引量:5
  • 2杨立信,徐志云,黄盛东,刘延玲,张宝仁.三种去细胞方法构建的猪主动脉瓣膜支架的性能比较[J].第二军医大学学报,2007,28(1):8-12. 被引量:8
  • 3Feldman T. Percutaneous valve repair and replacement: challenges encountered, challenges met, challenges ahead[J]. Circulation,2006,113:771-773.
  • 4Boudjemline Y, Agnoletti G, Bonnet D, Behr L, Borenstein N, Sidi D,et al, Steps toward the percutaneous replacement of atrioventricular valves an experimental study[J]. J Am Coll Cardiol, 2005,46 : 360-365.
  • 5Zegdi R, Khabbaz Z, Borenstein N, Fabiani J N. A repositionable valved stent for endovascular treatment of deteriorated bioprostheses[J]. J Am Coll Cardiol, 2006,48:1365-1368.
  • 6Feldman T, Martin B. Prospects for percutaneous valve therapies[J]. Circulation, 2007,116 : 2866-2877.
  • 7Block P C. Percutaneous transcatheter repair for mitral regurgitation[J]. J Interv Cardiol,2006,19:547-551.
  • 8Boudjemline Y,Agnoletti G, Bonnet D,Behr L, Borenstein N, Sidi D,et al. Steps toward the pereutaneous replaeement of atrioventrieular valves an experimental study[J]. J Am Coll Cardiol, 2005,46: 360-365.
  • 9Zegdi R,Khabbaz Z,Borenstein N, Fabiani J N. A repositionable valved stent for endovascular treatment of deteriorated bioprostheses[J]. J Am Coll Cardiol,2006,48 : 1365-1368.
  • 10Yacoub M H,Cohn L H. Novel approaches to cardiac valve repair from structure to function: Part Ⅱ [J]. Circulation, 2004, 109 : 1064-1072.

共引文献2

同被引文献47

  • 1Bonhoeffer P, Boudjemline Y, Saliba Z, et al. Transeatheter implantation of a bovine valve in pulmonary position ; a lamb study[J]. Circulation, 2000, 102(7): 813-816.
  • 2Bonhoeffer P, Boudjemline Y, Saliba Z, et al. Percutaneous replacement of pulmonary valve in a right-ventricle to pulmonary-artery prosthetic conduit with valve dysfunction[J]. Lancet,2000,356(9239) : 1403-1405.
  • 3Bouzas B, Kilner PJ , Gatzoulis MA. Pulmonary regurgitation: not a benign lesion[J]. Eur Heart J, 2005, 26 (5) : 433-439.
  • 4Discigil B, Dearani JA, Puga FJ, et al. Late pulmonary valve replacement after repair of tetralogy of Fallot[J].J Thorac Cardiovasc Surg, 2001,121(2)344-351.
  • 5Champsaur G, Robin J, Curtil A, et al. Long-term clinical and hemodynamic evaluation of porcine valved conduits implanted from the right ventricle to the pulmonary artery[J]. J Thorac Cardioxrasc Surg, 1998,116 (5) :793-804.
  • 6Lurz P, Coats L, Khambadkone S, et al. Percutaneous pulmonary valve implantation: impact of evolving technology and learning curve on clinical outcome[J]. Circulation, 2008, 117(15) : 1964-1972.
  • 7Andersen HR,Knudsen LL,Hasenkam JM.Transluminal implantation of artifical heart valve:description of a new expandable aortic valve and initial results with implantation by catheter technique in closed chest pigs.Eur Heart J,1992,13:704-708.
  • 8Cribier A,Ehchaninoff H,Bash A,et al.Percutaneous transcatheter implantation of an aortic valve prosthesis for calcific aortic stenosis:first human case description.Circulation,2002,106:3006-3008.
  • 9Grube E,Schuler G,Buellesfeld L,et al.Pereutaneous aortic valve replacement for severe aortic stenosis in high-risk patients using the second and current third generation self expanding CoreValve prosthesis.J Am Coll Cardiol,2007,50:69 76.
  • 10Piazza N,Grube E,Gerckens U,et al.Procedural and 30-day outcomes following transcatheter aortic valve implantation using the third generation (18 Fr) corevalve revalving system:results from the muhicentre,expanded evaluation registry 1-year following CE mark approval.Eurointervention,2008,4:242-249.

引证文献6

二级引证文献91

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部