摘要
本文描述用于放射性次级束终端中作△E测量的大面积(500mm×500mm)阳极分条读出双叠层电离室的构造和性能。65kPa的P10气体,对α源(8.785MeV)的能量分辨为2.5%,并用△E-TOF方法得到了较好的同位素鉴别。
This paper described the structure and performance of a large area ( 500 mm ×500 mm ) twin ionization chamber for the measurement of ΔE on radioactive second beam line. The energy resolution is 2.5% (α source 8.785 MeV ) when the P10 gas pressure is 65kPa. A better isotope identification has been achieved using ΔE TOF method. (
出处
《核电子学与探测技术》
CAS
CSCD
北大核心
1998年第2期136-138,153,共4页
Nuclear Electronics & Detection Technology
关键词
双叠层
同位素鉴别
电离室
核次级束
Large area Stripes of anode Twin fold Energy resolution Isotope identification)