期刊文献+

^6Li和^7Li双玻璃闪烁体γ补偿法测量中子注量 被引量:2

Neutron Flux Measurement with 6Li and 7Li Dual Glass Scintillators by γ Compensation Method
下载PDF
导出
摘要 依据6Li玻璃闪烁体对中子、γ皆灵敏,而7Li玻璃闪烁体仅对γ灵敏的特性,研究成功用6Li、7Li双玻璃闪烁体γ补偿法,实现在强γ辐射场干扰下,对弱中子注量率探测的方法。当60Coγ在中子信号幅度区的干扰计数率,占中子计数率的18.7%时,该方法对净中子注量测量结果导致的误差约为1%。 Based on the caracteristics of 6Li glass scintillator which is sensitive to both neutron and gamma rays, and 7Li glass scintillator which is sensitive to gamma rays only, a new method of detecting weak neutron flux under interference of strong gamma radiation has been investigated by mans of 6Li 7Li dual glass scintillator gamma compensation method. The result of neutron flux measurement by above mentioned method with an error of about 1% when the gamma ray interferenec is up to 18 7% has been obtained. (
机构地区 北京核仪器厂
出处 《核电子学与探测技术》 EI CAS CSCD 北大核心 1998年第3期182-186,共5页 Nuclear Electronics & Detection Technology
关键词 玻璃闪烁体 中子-Γ甄别 锂同位素 中子注量 Neutron Glass scintillator n γ discrimination)
  • 相关文献

参考文献3

  • 1汲长松,中子探测实验方法,1998年
  • 2汲长松,核辐射探测器及其实验技术手册,1990年
  • 3汲长松,核电子学与探测技术,1983年,13卷,2期,62页

同被引文献14

  • 1郭士伦.用塑料径迹探测器直接测量快中子的初步探讨[C].射线物理测量方法专业会议资料汇编[A].北京:原子能出版社,1975.
  • 2Chu K.Fenster A.Determination of neutron spectral including characteristic line intensitiens from transmission measuments using K-edge filter[J].Med phys,1983.(10)6:772~777.
  • 3Fleshier.The theory and application of solid nuclear tracks method[M].Stanford University Press,1979.
  • 4Lee H Y, Taddeucci T N, Haight R C, et al. Li-glass detector response study with a 252Cf source for low-energy prompt fission neutrons[J]. Nuclear Instruments and Methods in Physics Research A, 2013, 703:213-219.
  • 5Smith F P, Lewin D J. Dark matter detection[J]. Physics Reports, 2002, 187(5): 203-280.
  • 6Ryzhikov V, Nagornaya L, Volkov V, et al. Thermal neutron detectors based on complex oxide crystals[J]. Nuclear Instruments and Methods in Physics Research A, 2002, 486:156-159.
  • 7Booth T E, Goorley J T, Sood A, et al. MCNP5 users guide[R], http://www.lanl.gov/index.php, 2003.
  • 8Agostinelli S, Allision J, Amako K, et al. Geant4-a simulation toolkit[J]. Nuclear Instruments and Methods in Physics Research A, 2003, 506:250-303.
  • 9Mares V, Schraube H. Evaluation of the response matrix of a Bonner sphere spectrometer with LiI detector from thermal energy to 100 MeV[J]. Nuclear Instruments and Methods in Physics Research A, 1994, 337:461-473.
  • 10苏明,仲启平,郑玉来,张奇玮,阮锡超,周祖英,孔祥忠.γ全吸收型探测装置中子束流监视器的Geant4模拟[J].原子能科学技术,2009,43(10):946-950. 被引量:7