潮气量设置对部分二氧化碳重复吸入心输出量测定准确性影响的动物实验研究 被引量:2

Effect of tidal volume setting on the accuracy of partial CO_2 rebreathing technique for cardiac output monitoring by animal experiment
下载PDF
导出
摘要 目的:探讨机械通气潮气量设置对部分CO2重复吸入心输出量测定准确性的影响,为临床应用提供依据。方法:选用重庆地区健康雄性杂种犬8只,麻醉后气管切开机械通气。接无创心输出量监测仪(美国NICO),应用部分CO2重复吸入法测定心输出量(CORB);同时经右颈内静脉插入肺动脉漂浮导管,以热稀释法测定有创心输出量(COTD)。改变机械通气参数潮气量设置,分别在潮气量12ml/kg、6ml/kg条件下用两种方法同时测定心输出量,两种方法所测心输出量进行自体配对相关性研究。结果:配对t检验示潮气量为12ml/kg时,CORB为5.40 L/min±1.35 L/min,COTD为5.31 L/min±1.42 L/min,CORB和COTD比较,差异无显著性(P>0.05);CORB和COTD的偏差、精确性分别为0.19和1.06,相关系数为0.90,容许范围为-1.78 L/min至+2.14L/min;Vt设置为6ml/kg时CORB为4.21 L/min±1.20L/mln,COTD为5.65 L/min±1.61 L/min,CORB较COTD低,差异有显著性(P<0.05),两者的偏差、精确性分别为-1.68和1.04,相关系数为0.74,容许范围为-3.78 L/min至+1.58L/min。结论:部分二氧化碳重复吸入无创心输出量与热稀释有创心输出量相关性良好,潮气量是影响CORB准确性的因素之一。 Objective To explore the Effect of tidal volume setting on the accuracy of partial CO2 rebreathing technique for cardiac output monitoring in order to provide the evidence of clinical application. Methods Eight healthy anesthetized and ventilated dog were enrolled in this study. Each animal was inserted with Swan - Ganz catheter , and was installed a partial CO2 rebreathing ventilator circuit (Novametrix Medical SysteMs Inc, NICO). Simultaneous measurement of CORB, and CORD, were performed under two different ventilatory settings:(1) volume controlled ventilation with tidal volume of 12ml/kg(VCV, VT 12ml/kg) ;(2)volume controlled ventilation with tidal volume of 6ml/kg( VCV, VT 6ml/kg) ; Paired t - test and regression statistics were used to compare the degree of agreement between concurrent measurement of COn, and CORb. Results Paired t - test analysis revealed no significant difference between COW and COTD measurement when VCV at VT 12ml/kg( P 〉 0. 05 ). Under this ventilator), setting, Regression analysis revealed an R value of 0.90 and the bias and limits of agreement were as follows : 0. 19 and - 1.78L/min - + 2.14L/min; while,there is also a significant correlation was found between CORB and COrn when Vt setting at 6ml/kg: R = 0.74, bias - 1.68, limits of agreement -3.78L/min - + 1.58 L/min, the CORs, technique underestimated cardiac, output compared with COrn technique(P 〈 0. 05). Conclusion There is a high correlation between CORB, and COrD measurement. Tidal volume is one of the important factor affecting the accuracy of CORB.
出处 《激光杂志》 CAS CSCD 北大核心 2009年第1期90-91,共2页 Laser Journal
基金 重庆医科大学附属第一医院启动基金(2000-12)
关键词 心输出量 部分二氧化碳重复吸入 热稀释法 cardiac output tidal vohune partial CO2 rebrcathing thermodilution
  • 相关文献

参考文献8

  • 1Horst H M, Obeid FN, ViJ D, et al. The risks of pulmonary arterial catheriztion[J]. Surg 蒙古 Gynecol Obster, 1984,159:229 - 232.
  • 2Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement[J]. Lancet 1986, 1 : 307 - 310.
  • 3Bland MJ, Altman DG. Calculating correlation coefficients with repeated observation: Part Ⅰ - Correlation within subjects[J]. Br Med J, 1995, 310-446.
  • 4Binder JC, Parkin WG. Non - invasive cardiac output determination: comparison of a partial - rebreathing technique with thennodilution[ J]. Anaesth Intensive Care, 2001,29( 1 ) : 19 - 23.
  • 5刘琼,黄纪坚.部分二氧化碳重复吸入法与温度稀释法测定心排血量的相关性研究[J].临床麻醉学杂志,2003,19(7):390-391. 被引量:2
  • 6刘琼,黄纪坚.二氧化碳重复吸入无创心排血量与热稀释有创心排血量相关性动物实验对比研究[J].重庆医科大学学报,2007,32(11):1182-1184. 被引量:2
  • 7Neuhauser C, Muller M, Brau M, et al. Partial CO2 rebreathing technique versus thermodilution: measurement of cardiac output before and after operation extracorporeal circulation [ J ]. Anesthesiology, 2002, 51 (8) : 625 - 633.
  • 8Odenstedt H, Stenqvist O, Lundin S. Clinical evaluation of a partial CO2 rebreathing tectmique for cardiac output monitoring in critically ill patients[J]. Acta Anaesthesiol Scand,2002,46: 152.

二级参考文献14

  • 1Osterlund B,Gedeon A,Krill P,et al. A new method of using gas exchange measurement for the noninvasive determination of cardiac output, clinical experiences in adults following cardiac surgery. Acta Anaesthesiol Scand,1995,39:727.
  • 2Bailey PL, Haryadi DG, Orr JA, et al. Partial CO2 rebreathing Fick technique for noninvasive measurement of cardiac output. Anesthesia Analgesia, 1998,86:124.
  • 3Watt RC, Loeb RG,Orr J. Comparison of a new non-invasive cardiac output xeehnique with invasive bolus and continuous thermodilution. Anesthesiology, 1998,89 : 536.
  • 4Maxwell RA,Gibson JB,Slade JB,et al. Noninvasive cardiac output by partial CO2 rebreathing after severe chest traum. Traum,2001,51.849.
  • 5Odenstedt H, Stenqvist O, Lundin S. Clinical evaluation of a partial CO2 rebreathing technique for cardiac output monitoring in critically ill patients, Acta Anaesthesiol Scand,2002,46 : 152.
  • 6Johnson KB, Haryadi DG,JA Orr,et al. Influence of pulmonary edema on noninvasive measurement of cardiac output using partial CO2 rehreathing in a canine model.Anesthesiology, 1998,89 : 535.
  • 7Horst H M,Obeid FN,ViJ D,et al. The risks of pulmonary arterial catheriztion[J]. Surg Gynecol Obster, 1984; 159:229.
  • 8Blomqvist H,Frostell C,Hedenstriema G. A Non-invasive technique for measurement of lung perfusion[J]. Intensive Care Med, 1986; 12 : 172.
  • 9Kuck K,Orr JA,Haryadi DG,et al. Evaluation of the NICO2 partial co2 rebreathing cardiac output monitor in spontaneously breathing animals[J]. Anesthesiology, 1999 ;91 (3A) : A560.
  • 10Binder JC,Parkin WG. Non-invasive cardiac output determination:comparison of a partial-rebreathing technique with thermodilution[J]. Anaesth Intensive Care,2001 ;29( 1 ) : 19-23.

共引文献2

同被引文献22

  • 1华田苗,梅斌,王海涛,刘再群,孙庆艳.特定方位的光栅识别学习不改变猫外膝体背核神经元的方位敏感性[J].动物学报,2005,51(2):274-279. 被引量:1
  • 2姜威,王琳,楚东岭,金发光.不同潮气量机械通气过程中呼吸机相关性肺损伤的病理学变化[J].中国组织工程研究与临床康复,2007,11(31):6236-6238. 被引量:14
  • 3Mathews L,Singh RK.Cardiac output monitoring.Ann Card Anaesth,2008,11(1):56-68.
  • 4Ng JM,Chow MY,Ip-Yam PC,et al.Evaluation of partial carbon dioxide rebreathing cardiac output measurement during thoracic surgery.J Cardiothorac Vasc Anesth,2007,21(5):655-658.
  • 5Carretero MJ,Fontanals J,Agustí M,et al.Monitoring in resuscitation:comparison of cardiac output measurement between pulmonary artery catheter and NICO.Resuscitation,2010,81(4):404-409.
  • 6Haryadi DG,Orr JA,Kuck K,et al.Partial CO2 rebreathing indirect Fick technique for non-invasive measurement of cardiac output.J Clin Monit Comput,2000,16(5-6):361-374.
  • 7Tachibana K,Imanaka H,Takeuchi M,et al.Effects of reduced rebreathing time,in spontaneously breathing patients,on respiratory effort and accuracy in cardiac output measurement when using a partial carbon dioxide rebreathing technique:a prospective observational study.Crit Care,2005,9(5):R569-574.
  • 8Allardet-Servent J,Forel JM,Roch A,et al.Pulmonary capillary blood flow and cardiac output measurement by partial carbon dioxide rebreathing in patients with acute respiratory distress syndrome receiving lung protective ventilation.Anesthesiology,2009,111(5):1085-1092.
  • 9Guret G,Rossignol B,Kiss G,et al.Cardiac output measurements in off-pump coronary surgery:compareison between NICO and the Swan-Ganz catheter.J Euro Anesthesiology,2006,23:848-854.
  • 10Kalairajah Y,Sinpson D,Cossey AJ,et al.Blood loss after total knee replacement effects of computer-assisted surgery.J Bone Joint Surg Br,2005,87(11):1480-1482.

引证文献2

二级引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部