摘要
在pH 5.5的Na2HPO4-柠檬酸缓冲溶液中,吖啶橙(AO)与PAR-钒(Ⅴ)络合物之间发生有效的能量转移,使吖啶橙526 nm处的荧光猝灭,据此建立了能量转移荧光测定痕量钒的新方法。实验表明,在pH5.5的3.0×10^-5mol.L^-1AO-4.0 mg.L^-1PAR体系中,钒的含量c在0.12-0.5μg.mL^-1范围内与荧光猝灭强度ΔF呈良好线性关系,线性回归方程为ΔF=165.4c+2.5,方法检出限为0.0045μg.mL^-1,相对标准偏差为0.6%。该荧光猝灭反应15 min内完全,其相对荧光强度在2.5 h内基本保持不变。用该方法测定生物样品中钒的含量,相对误差小于5%,结果满足痕量分析要求。
An energy transfer technique for 4-(2-pyridylazo) resorcinol (PAR)-vanadium( V)and acridine orange(AO)was studied, and the optimum conditions of energy transfer system were also experimented. It was found that in eitrate-Na2HPO4 buffer solution at pH=5.5, energy transfers from AO to vanadium( V )-PAR complexes. A new method based on energy transfer fluorescence quenching for the determination of trace vanadium( V )with AO-PAR-V( V )was established. The equation of linear regression is AF=165. 4c+2. 5, and the determination range of vanadium is 0. 012-0. 5 ug · mL^-1 , with detection limit of 0. 004 5 ug · mL^-1. The correlation coefficient is R=0. 998 5, and relative standard deviation is 0. 6%. The fluorescence reaction is completed within 15 minutes, and relative fluorescence intensity remains unchanged for 2.5 hours. The influence of foreign ions on the determination of V (0. 5 9g · mL^- 1 ) was examined, with a related error of 5%. The method has been used in the determination of trace vanadium in biological samples with the relative error of 6. 98%, which meets the requirements of trace analysis.
出处
《光谱学与光谱分析》
SCIE
EI
CAS
CSCD
北大核心
2009年第2期476-478,共3页
Spectroscopy and Spectral Analysis
基金
国家自然科学基金资助项目(20365001)
广西自然科学基金资助项目(桂科自0728218,0679005)
桂林工学院科研启动基金项目
广西壮族自治区教育创新人才基地项目资助
关键词
吖啶橙
PAR
能量转移
荧光猝灭
钒
Acridine orange
4-(2-Pyridylazo) resorcinol( PAR)
Energy transfers
Fluorescence quenching
Vanadium