Pentacam三维眼前节分析仪测量中央角膜厚度 被引量:5

Measuring central corneal thickness with Pentacam Scheimpflug imaging system
下载PDF
导出
摘要 目的评价Pentacam三维眼前节分析仪测量中央角膜厚度的准确性。方法分别使用Pentacam三维眼前节分析仪和A超角膜测厚仪测量75例(150眼)门诊病人的中央角膜厚度,比较两种方法得到的测量结果,并做相关分析。两种方法测得值相差在±25μm之内者,认定为结果相符。结果Pentacam三维眼前节分析仪测得中央角膜厚度为(535.42±25.27)μm,A超角膜测厚仪测得中央角膜厚度为(539.65±37.23)μm,两种测量结果间比较差异无统计学意义(P>0.05);经相关分析显示具有相关性(P<0.01)。结果符合率为94.67%。结论Pentacam三维眼前节分析仪测量中央角膜厚度结果准确,有较好的临床应用前景。 Objective To assess the accuracy of Pentacam Scheimpflug imaging system in the measurement of central corneal thickness. Methods A-scan ultrasound pachymetry and Pentacam Scheimpflug imaging system were used to measure the central corneal thickness on 75 outpatients( 150 eyes). The results were compared. If the difference between these two methods was within -25 -25 um, the results were considered as coincidence. Results The value of central corneal thickness measured by the Pentacam Scheimpflug imaging system was (535.42 ± 25.27 ) um. The value of central corneal thickness measured by A-scan ultrasound pachymetry was (539.65 ±37.23) um. There was no significant difference between the results of there two methods(P 〉0.05). There was a significant linear correlation between the two methods(P 〈 0. 01 ). Coincidence rate was 94.67%. Conclusion The results obtained by Pentacam Scheimpflug imaging system are accurate, suggesting the Pentacam system a promising new instrument for central corneal thickness evaluation.
出处 《现代医学》 2009年第1期38-40,共3页 Modern Medical Journal
关键词 角膜 超声检查 Scheimpflug摄像法 corneal ultrasound pachymetry Scheimpflug images
  • 相关文献

参考文献9

  • 1Shah S, Chatterjee A, Mathai M, et al. Relationship between corneal thickness and measured intraocular pressure in a general ophthalmology clinic [ J ]. Ophthalmology, 1999,106 ( 11 ) : 2154-2160.
  • 2Marsich M W, Bullimore M A. The repeatability of corneal thickness measures [ J ]. Cornea,2000,19 (6) : 792- 795.
  • 3李坤,陆蓓,崇晓霞,陈彦玲,李晓玲.麻醉剂和泪膜对健康人角膜厚度的影响[J].眼视光学杂志,2004,6(4):244-246. 被引量:16
  • 4Kim S W, Byun Y J, Kim E K, et al. Central corneal thickness measurements in unoperated eyes and eyes after PRK for myopia using Pentacam, Orbscan Ⅱ , and ultrasonic pachymetry [ J ]. J Refract Surg, 2007,23 ( 9 ) : 888- 894.
  • 5Lackner B, Schmidinger G, Pieh S, et al. Repeatability and reproducibility of central corneal thickness measurement with Pentacam, Orbscan, and ultrasound [ J ]. Optom Vis Sci, 2005, 82(10) :892-899.
  • 6Ciolino J B, Khachikian S S, Belin M W. Comparison of corneal thickness measurements by ultrasound and scheimpflug photography in eyes that have undergone laser in situ keratomileusis [J]. Am J Ophthalmol,2008,145 (1) :75-80.
  • 7De Sanctis U, Missolungi A, Mutani B, et al. Reproducibility and repeatability of central corneal thickness measurement in keratoconus using the rotating Scheimpflug camera and ultrasound pachymetry [ J ]. Am J Ophthalmol, 2007, 144 ( 5 ) : 712-718.
  • 8Gardner B G, Taub M B. New diagnostic devices in Eyecare [ J]. Rev Optom,200d, 15:79-85.
  • 9Kent C. The anterior chamber from every angle [ J ]. Rev Ophthalmol,2005 ,12 :33-38.

二级参考文献9

  • 1Yaylali V,Kaufman SC,Thompsom HW. Corneal thickness measurements with the Orbscan Topography System and Ultrasound Pachymetry[J]. J Cataract Refract Surg,1997,23(9):1345-1350.
  • 2李凤鸣.眼科全书(上册)[M].北京:人民卫生出版社,1997.42.
  • 3Iskander NG,Anderson Penno E,Peter NT,et al. Accuracy of Orbscan pchymetry measurements and DGH ultrasound pachymetry in primary laser in situ Keratomileusis and LASIK enhancement procedures[J]. J Cataract Surg,2001,27(5):681-685.
  • 4Nissen J,Hjortdal JO,Ehlers N,et al. A clinical comparison of optical and ultrasonic pachymetry[J]. Acta Ophthalmol Copenh,1991,69(5):659-663.
  • 5Lattimore MR Jr. Influence of extended soft contact lens wear on the comparative measurenent of central corneal thickness[J]. Acta Ophthalmol Scand,1996,74(2):239-242.
  • 6Mertz GW. Overnight swelling of the living human cornea[J]. J Am Optom Assoc,1980,51(2):211-214.
  • 7Lattimore MR Jr,Sandor Kaupp MS,Schallhorn S,et al. Orbscan pachymetry:implication of a reperted measures and diurnal variation analysis[J]. Ophthalmology,1999,106(5):977-981.
  • 8Dursun D,Monroy D,Knighton R,et al. The effects of experimental tear film removed on corneal surface regularity and barrier function[J]. Ophthalmology,2002,107(9):1754-1760.
  • 9谢立信,李赵霞,胡隆基,衣伟伟.Orbscan-Ⅱ在准分子激光原位角膜磨镶术术前检查中的应用[J].眼视光学杂志,2001,3(3):133-136. 被引量:13

共引文献15

同被引文献104

引证文献5

二级引证文献39

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部