高强度聚焦超声辐照剂量与生物学效应研究 被引量:6

Study on Therapeutic Dosimetry and Biologic Effect of High Intensity Focused Ultrasound
原文传递
导出
摘要 研究高强度聚焦超声体外致凝固性坏死的辐照剂量及其对应产生的生物学效应。将新鲜离体牛肝组织(屠宰后6 h内),浸泡于0.9%NS中,常规脱气25 min后作为实验材料备用。实验设备为JC型高强度聚焦超声肿瘤治疗系统。治疗头频率1.6 MHz。采用定点点打方式辐照牛肝组织。固定辐照深度为20 mm。辐照声功率为65.44~196.32 W,分为8级,每级相差16.44 W,对应每级声功率,辐照时间分别为1~20 s。声功率与辐照时间的乘积即为辐照剂量。相同剂量下重复实验3次。取靶区灰白色区域与正常组织交界处组织送病理检查(HE染色),光镜下观察靶区和靶区边缘是否发生凝固性坏死。实验发现辐照所采用的声功率越大,达到凝固性坏死所需辐照时间越短,当声功率达到一定水平(超过179.96 W)只需1s就能致凝固性坏死;不同辐照剂量,生物学焦域形态不同,生物学焦域面积大小亦不同,生物学焦域面积与辐照剂量之间存在正相关(y=0.0164x1.0591,R2=0.9238,P<0.05)。因此,研究高强度聚焦超声辐照剂量与生物学效应关系,有助于临床治疗过程中辐照剂量的选择。 This study was aimed at exploring the high intensity focused ultrasound(HIFU) therapeutic dosimetry and its biologic effect. In-vitro, bovine liver was immersed into 0.9% NS and degassed for application. The JC Model-focused ultrasound tumor therapeutic system was used in the experiment. The HIFU parameters were: frequency 1.6 MHz,depth 20 mm, acoustic power from 16.44 W to 196.32W. Under each power, at radiating times from ls to 20 s, bioptic specimens were obtained from all samples. The results showed when the acoustic power was equal to or higher than 179.96 W, only ls of radiating is adequate to induce eoagulative necrosis, and when the acoustic power was lower than 65.44 W, the radiating time to produce coagulative necrosis was about 7 s. In the range of 65.44-179.96 W, at each time when the acoustic power was set up with an increment of 16.36 W, the time to produce coagulative necrosis was 1-2 s shorter. The form of biological focal region(BFR) varied with the acoustic power and HIFU irradiation time. The size of BFR increased with the increase of HIFU irradiation dosage (a- coustic power)〈 exposure time). There is positive correlation between the size of BFR and the dosage of HIFU irradi- ation(y=0.0164x1.0591,R2=0.9238,P〈O.05).
出处 《生物医学工程学杂志》 EI CAS CSCD 北大核心 2009年第1期72-74,84,共4页 Journal of Biomedical Engineering
关键词 高强度聚焦超声 辐照剂量 生物学效应 High intensity focused ultrasound(HIFU) Therapeutic dosimetry Biologic effect
  • 相关文献

参考文献6

  • 1李发琪,王智彪,杜永洪,马平,白晋,伍烽,冯若.高强度聚焦超声“切除”组织的剂量学研究[J].生物医学工程学杂志,2006,23(4):839-843. 被引量:34
  • 2Ter H, Sinnett D GR, I Rivens. High intensity focused ultrasound :A surgical technique for the treatment of discrete liver tumors. Phys Med Biol,1989;34 : 1743
  • 3Vaezy S, Andrew M, Kaczkowski P, et al. Image-guided acousticthrapy. Annu Rev Biomed Eng, 2001;3 : 375
  • 4Fry FJ,Reitly CR,Dines KA,etal. Absorption in liver at the focus of an ultrasonic shock wave field. Ultrasound Med Biol, 1991;17: 65P
  • 5伍烽,王智彪,陈文直,白晋,邹建中,王芷龙,阮祥燕,杜永洪,孔繁兵,朱辉.高强度聚焦超声体外治疗肝肿瘤的剂量学研究[J].中华物理医学与康复杂志,2000,22(5):267-269. 被引量:22
  • 6Wang ZB, Bai J, Li FQ,et al. Study of a biological focal region of high-intensity focused ultrasound. Ultrasound in Medicine & Biology,2003,29 :49

二级参考文献15

  • 1Sibille A, Prat F, Chapelon JY, et al. Characterization of extracorporeal ablation of normal and tumor-bearing liver tissue by high intensity focused ultrasound. Ultrasound Med Biol, 1993, 19(8) : 803
  • 2Adams JB, Moore RG, Anderson JH, et al. High-intensity focused ultrasound ablation of rabbit kidney tumors. J Endourol, 1996, 10 : 71
  • 3Bihrle R, Foster RS, Sanghvin NT, et al. High intensity focused ultrasound for treatment of benign prostatic hyperplasia: early United States clinical experience. J Urol,1994,151:1271
  • 4Malcolm AL, ter Haar GR. Ablation of tissue volumes using high intensity focused ultrasound. Ultradsound Med Biol,1996,22(5) : 659
  • 5ter Haar GR, Robertson D. Tissue destruction with focused ultrasound in vivo. Eur Urol, 1993,23(suppl 1) : 8
  • 6Wang ZB, Bai J,Li FQ, et al. Study of a “biological focal region”of high-intensity focused ultrasound. Ultrasound Med Biol, 2003, 29(5) : 749
  • 7Yang R, Sanghvi NT, Rescorla FJ,et al. Liver cancer ablation with extracorporeal high-intensity focused ultrasound. Eur Urol, 1993,23 (Suppl. 1) :17
  • 8Chapelon JY, Margonari J, Theillere Y, et al. Effects of high-intensity focused ultrasound on kidney tissue in the rat and the dog. Eur Urol, 1992,22 :147
  • 9Chen L, ter Haar G, Hill CR. Influence of ablated tissue on the formation of high-intensity focused ultrasound lesions.Ultrasound Med Biol, 1997, 23(6) : 921
  • 10Chen L, Rivens I, ter Haar GR. , et al . Histological change in rat liver tumor treated with intensity focused ultrasound.Ultrasound Med Biol, 1993, 19 :67

共引文献53

同被引文献115

引证文献6

二级引证文献25

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部