400mgI/ml和370mgI/ml造影剂在脊髓血管CTA中的应用对比 被引量:1

Comparison of effectiveness of 400 mgI/ml and 370 mgI/ml nonionic contrast medium on spinal CTA
下载PDF
导出
摘要 目的:比较400 mgI/ml与370 mgI/ml非离子碘造影剂在脊髓血管CTA中能否提高脊髓前动脉、大根髓动脉的图像质量。方法:103例患者分为370 mgI/ml组(51例)和400 mgI/ml组(52例),行多排CT脊髓血管CTA扫描,图像经颅脑CT算法及小视野二次重建,对比(1)降主动脉起始段CT值;(2)脊髓前动脉(ASA)和大根髓动脉(AA)显示率;(3)AA走行各段对比噪声比(CNR);(4)ASA和AA图像质量评分。结果:370 mgI/ml组和400 mgI/ml组降主动脉起始段血管内密度分别为(305.5±23.7)HU和(367.4±19.8)HU,组间有统计学差异(t=2.23,P<0.05);370 mgI/ml组和400 mgI/ml组的ASA和AA的显示率均为100%;370 mgI/ml组和400 mgI/ml组AA各段的CNR分别为:(1)AA起源血管段(肋间动脉或腰动脉):24.11±4.43和27.88±5.12,组间有统计学差异(t=3.998,P<0.05);(2)椎间孔段:7.23±4.28和7.87±4.72,组间无统计学差异(t=0.316,P>0.05);(3)椎管内段:6.08±1.17和6.53±1.52,组间无统计学差异(t=0.894,P>0.05);(4)汇入ASA动脉后:9.72±1.91和9.49±2.01,组间无统计学差异(t=0.564,P>0.05)。ASA和AA的图像质量评分显示370 mgI/ml组和400 mgI/ml组组间没有统计学差异(χ2=0.317,P>0.05;χ2=0.569,P>0.05)。结论:多排CT脊髓CTA中,使用400 mgI/ml非离子碘造影剂与370 mgI/ml非离子碘造影剂相比可以增加血管强化,但并不能显著提高ASA和AA图像质量。 Objective: To test the hypothesis that the 400 mgI/ml nonionic contrast medium can promote CTA image quality of the anterior spinal artery(ASA) and artery of Adamkiewicz ( AA ), compared with 370 mgI/ml nonionic contrast medium. Methods : 103 patients who underwent spinal CTA scan were divided into two groups : 370 mgI/ml group ( n = 51 ) and 400 mgI/ml group ( n = 52 ). Raw image data was reconstructed by adapted brain algorithms and small DFOV. It was studied : ( 1 ) CT value in originated part of the descendent aorta; ( 2 ) the visualization of ASA and AA in CTA; (3) contrast - to - noise ratios (CNR) at four deferent levels along the course of AA; (4) the image quality scores of ASA and AA. Results: CT value in originated part of the descendent aorta was (305.5 ±23.7 ) HU in 370 mgI/ml group and (367.4± 19.8 ) HU in 400 mgI/ml group, there was a statistic difference between them ( t = 2.23, P 〈 0.05 ). The ASA and AA can be 100% visualized in all patients of both groups. The CNR of the feeding artery of AA ( intercostal artery or lumber artery) was 24.11 ± 4.43 and 27.88 ±5.12 in 370 mgI/ml group and 400 mgI/ml group, respectively. There was a statistic difference between them ( t = 3. 998, P 〈 0.05 ). The CNR at the level of AA in vertebrae foramen, vertebrae canal and the cranial part of the ASA were 7.23 ± 4.28,6.08 ± 1.17 and 9.72 ± 1.91 in 370 mgI/ml group,as well as 7.87± 4.72,6.53 ± 1.52 and 9.49 ±2.01 in 400 mgI/ml group. However, all of CNR at the levels of the AA in vertebrae foramen ( t = 0. 316) , in vertebrae canal (t =0. 894) and the cranial part of the ASA( t =0. 564) did not show significant difference between the two groups (P 〉0.05 ). Furthermore, the image quality scores of ASA and AA indicated there were also no differences between them(Χ^2 = 0.317, P 〉 0.05 ;Χ^2 = 0. 569, P 〉0.05). Conclusion: Compared with 370 mgI/ml nonionic contrast medium,400 mgI/ml contrast medium can provide higher attenuation of blood vessels due to its higher iodine concentrations, but it can not promote the image quality of ASA and AA in multi - detector CT (MDCT) spinal CTA examination.
出处 《西南国防医药》 CAS 2009年第3期281-284,共4页 Medical Journal of National Defending Forces in Southwest China
关键词 脊髓血管造影 多排CT 造影剂 碘浓度 spinal angiography multi - detector CT contrast medium iodine concentration
  • 相关文献

参考文献11

  • 1SAVADER S J,WILLIAMS G M,TREROTOLA S O,et al. Preoperative spinal artery localization and its relationship to postoperative neurelogic complications [ J ]. Radiology,1993,189 ( 1 ) : 165 - 171.
  • 2CADEMARTIRI F, MOLLET N R, LEMOS P A, et al . Higher intracoronary attenuati on improves diagnostic accuracy in MDCT coronary angiography[ J]. A J R,2006,187(4) :W430 - W433.
  • 3BOLL D T,LEWIN J S,FLEITER T R,et al. Multidetector CT angio - graphy of arterial inflow and runoff in the lower extremities : a challenge in data acquisition and evaluation [ J ]. J Endovasc Ther,2004,11 (2) : 144 - 151.
  • 4YOSHIOKA K,NIINUMA H,OHIRA A,et al. MR angiography and CT angiography of the artery of Adamkiewicz: noninvasive preo - perative assessment of thoracoabdominal aorta aneurysm [ J ]. Radiographics ,2003,23 ( 5 ) : 1215 - 1225.
  • 5FLESICHMANN D. Use of high concentration contrast media:prin - ciples and rationale - vascular district [ J ]. Eur J Radiol, 2003,45 (Supp l1):S88 - S93.
  • 6KONIG M, BULTMANN E, BODE - SCHNURBUS L, et al. Image quality in CT perfusion imaging of the brain: the role of iodine concentration[ J ]. Eur Radiol,2007,17 ( 1 ) :39 - 47.
  • 7SCHOELLNAST H, DEUT SCHMANN H A, FRITZ G A, et al. MDCT angiography of the pulmonary arteries: influence of iodine flow concentration on vessel attenuation and visualization [ J ]. AJR, 2005,184(8) :1935 - 1939.
  • 8CADEMARTIRI F, MONYE C D, PUGLIESE F, et al. High iodine concentration contrast material for noninvasive muhislice computed tomography coronary angiography iopromide 370 versus iomeprol 400 [ J ]. Invest Radio1,2006,41 ( 3 ) :349 - 353.
  • 9SETTY B N, SAHANI D V, OUELLETTE - PIAZZO K, et al. Comparison of enhancement,image quality ,cost ,and adverse reactions using 2 different contrast medium concentrations for routine chest CT on 16 - slice MDCT[J]. J Comput Assist Tomogr,2006,30(5) :818 - 822.
  • 10李云生,李崇谦,王琨,陈仲欣,贺能树,王永赞.选择性脊髓动脉造影的应用解剖[J].中国临床解剖学杂志,1995,13(1):14-17. 被引量:7

二级参考文献5

共引文献6

同被引文献8

引证文献1

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部